Status Register System
2500A Series Operation Manual, 34172 Revision C, March 2008
condition has been enabled by the Standard Event Status Enable Register, bit 5 is set in the Status Byte
Register. The *ESR? query can be used to read the contents of the Standard Event Status Register.
The Standard Event Status Enable Register controls which bits in the Standard Event Status Register can
set bit 5 of the Status Byte Register. The bits in the Standard Event Status Enable Register are logically
ANDed with the equivalent bits in the Standard Event Status Register, and the results of those AND
operations are logically ORed to produce a summary bit. The ESB bit (bit 5) in the Status Byte Register
is set when the logic OR operation sets the summary bit. The *ESE command can be used to set the con-
tents of the Standard Event Status Enable Register, and the *ESE? query can be used to read the contents
of the Standard Event Status Enable Register.
The following describes each bit in the Standard Event Status Register:
0
Operation Complete. This bit is set whenever all pending operations are completed (such as a
list computation).
1
Not used. Always 0.
2
Query Error. This bit is set whenever a query error has occurred.
3
Not used. Always 0.
4
Execution Error. This bit is set whenever an execution error has occurred.
5
Command Error. This bit is set whenever an invalid GPIB command has been received.
6
Not used. Always 0.
7
Power On Occurrence. This bit is set whenever the instrument has been powered off and then on
again during manual and remote operation.
3.5.3
Questionable Status Condition and Enable Registers
The Questionable Status Register is one of the lower-level status registers. It monitors certain 2500A-
specific status conditions. When a condition occurs that is being monitored by this register, bit 3 is set in
the Status Byte Register. The STATus:QUEStionable:CONDition? query can be used to read the contents
of the Questionable Status Register.
7
6
5
4
3
2
1
0
Power on
Occurrence
Not Used
Command
Error
Execution
Error
Not Used
Query
Error
Not Used
Operation
Complete