5.
If the UUT includes the 1 Hz resolution option, perform this test. This test exercises the
1 Hz through 100 kHz digits in such a manner that any incorrect digital programming
data being sent to the 1 Hz Resolution assembly will be detected. Improper operation of
the 1 Hz assembly will also be indicated. Refer to the Test Data Sheet and program each
of the listed frequencies into the UUT.
Calibration and Testing
Manual No. 120AM00250, Rev C, September 1998
4-13