Gill Instruments Ltd
Intrinsically Safe WindObserver Anemometer
Page 43
Issue 10
Doc. No. 1360-PS-0001
May 2015
Returning to Measurement mode
Type Q and press ENTER key
If in
Continuous
mode, the anemometer responds with wind measurements immediately,
continuing at the selected Sampling rate.
If in
Polled
mode:-
?
Enables poll
N
Polls anemometer (where N is the Unit identifier entered as upper-case)
The anemometer replies with a single set of wind measurements
&
Anemometer replies with Unit identifier
!
Disables poll
Note:- If the unit is in Polled Mode it is always best to interrogate the unit for data with a
?
before the poll identifier to cater for the event that the power has been switched off or
power interrupted.
Checking the configuration
We strongly recommend that, as a standard procedure, you use this command (
D3
) prior
to, and after, changing any settings. It shows the current settings for all the alterable
settings. We suggest you note down your settings, so that you can easily return to them.
Type *
Enters Configuration Mode (from Continuous mode)
Or
Type *N
Enters Configuration Mode (from Polled mode)
Type D3 and press ENTER key
The Intrinsically Safe WindObserver responds with
the current configuration settings.
The factory default settings are:-
B3 F1 G000 K1 L1 M2 NA O1 P1 U1 V1 X1
To return to Measurement mode
Type Q and press ENTER key
How to change these settings is explained in the following sections.