Gill Instruments Ltd
Intrinsically Safe WindObserver Anemometer
Page 50
Issue 10
Doc. No. 1360-PS-0001
May 2015
Change settings if required referring to the previous configuration details.
Type Q and press Enter to go back into measurement mode.
If powering down the instrument and repowering and no change has been made to the
baud rate (B command ) then open a new terminal program at the units original baud rate
setting (the default setting would be 9600 Bauds to view data).
10.5 Status (error) codes
The Status code is sent as part of each wind measurement message.
Code
Status
Condition
00
OK
Sufficient samples in average period
A
OK
NMEA data Acceptable
01
Axis 1 failed
Insufficient samples in average period on U axis
02
Axis 2 failed
Insufficient samples in average period on V axis
04
Axis 1 and 2 failed
Insufficient samples in average period on both axes
08
NVM error
NVM checksum failed, data could be uncalibrated.
09
ROM error
ROM checksum failed, data could be uncalibrated.
51
Measurement average
building.
Data valid but warns that average period not reached
when averaging used.
V
NMEA data Void
Invalid data output