Gill Instruments Ltd
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
WindObserver II Pipe Mount
Page 34
Issue 04
Doc. No. 1390-PS-0009
August 2009
Vx - Vertical Output Padding
Setting
Configuration code
Disable vertical output padding
V1
Enable vertical output padding
V2
Xx - 45
Alignment Offset/Inverted Operation.
Setting
Configuration
code
Notes
Align U axis with
transducer axis
X1
X1. Aligns U axis with North/South axis.
Align U axis +45
to transducer axis
X2
X2. This re-aligns both U & V and polar by
45
˚.
Reverses Polar
Direction
X3
X3 reverses reported polar direction to allow the
instrument to be mounted upside down. N.B.
Does NOT affect UV alignment (Mode 1, 3).
Align @ 45° from
North
X4
X4 set polar alignment at 45 degrees from North
when instrument is mounted upside down. Does
not affect UV alignment (Mode 1,3).
Figure 4 shows the polarity of U and V if the wind components along the U and V axis are
blowing in the direction of the respective arrows.
-U
+U
+V
-V
E
N
S
W
Figure 4 UV Polarity
Yx - Not Applicable.
Zx – Not Applicable.
陕西威瑞仪器仪表有限公司 生产代理销售:实验检测设备气象科学仪器设备