WindMaster and WindMaster Pro
1561-PS-0001.
Issue 11 February 2017
________________________________________________________________________________________________________________________
53
Request Current Configuration
D3
M2,U1,O1,L1,P1,B4,H1,NQ,E1,T1,S1,C2,A1,I1,J1,V1,X1,G0,K50
Report DAC Channel Configuration.
D7
DAC CHAN 01, MODE 05, CONFIG 02
DAC CHAN 02, MODE 09, CONFIG 01
DAC CHAN 03, MODE 04, CONFIG 01
DAC CHAN 04, MODE 08, CONFIG 01
Where:-
Mode 01 = DAC Channel Off.
Mode 02 = U Magnitude Output.
Mode 03 = V Magnitude Output.
Mode 04 = W Magnitude Output.
Mode 05 = Polar Angle Output.
Mode 08 = Status (Good/Bad).
Mode 09 = Polar Magnitude Output.
and Configuration code for Mode 08 equates to good/bad status setting.
Config 01 = Good is high output level, Bad is low level.
Config 02 = Good is low output level, Bad is high level.
and Configuration codes for Modes 02, 03, 04, and 09 equate to scale setting:-
Config 01 = 5m/s
to
Config 08 = 120m/s
and Configuration code for Mode 5 equates to angle wrapping where:-
Config 01 = 540 degree wraparound.
Config 02 = 360 degree wraparound.
Ex Set Physical Communications
1
AUTO
2
485
3
232
It is recommended to leave the unit in AUTO mode. This can also be overridden from RS232 to RS422/485 by:-
On 31 way connector units without a Pipe Mount Box link the comms line to V+ for RS232 or to 0V for RS422/485.
On units fitted with a Pipe mount box if the link in the box is in its default position (see page 19) then the above
external link connections apply.
On a unit fitted with a Pipe Mount box the internal link in the box may be moved to set R232 and RS422/RS485
operation (see Page 19). If the link is moved from its default position then an external wire link should not be used.
See section 6.6.1 and 6.6.2.
If a unit is set for E3 and required to be set back to E1 then with a HyperTerminal connection established go into
Configuration Mode, Type E1 and Press Enter. Then Type E and Press Enter. Type Q and press Enter to go back into
measurement mode.