WindMaster and WindMaster Pro
1561-PS-0001.
Issue 11 February 2017
________________________________________________________________________________________________________________________
54
Fx y Signal Processing
Retries Enabled.
WindMaster Default setting is ON.
Enabled – If this mode is selected, if a problem sample is detected, the unit will attempt another ultrasonic firing within
the set time parameters.
Disabled – The unit will not attempt to re-check problem samples.
11.
Instantaneous Sampling
Default is OFF.
Enabled – The WindMaster can be configured to output the sampled data direct at the selected output rate without any
averaging -"Instantaneous Sampling". If for example the sensor is set for 1Hz output rate then the unit will undertake
one set of transducer firings and output the result after the 1 second period.
Disabled – Alternatively data can be sampled at 20Hz or 32Hz and averaged. This applies for wind data and speed of
sound. All samples are transmitted with a status code.
If the sensor is for example set to 1 Hz output rate then the unit will undertake a full set of firings (at 20 Hz or 32Hz)
and output the average result of them after the 1-second period.
12.
Calibration Enabled
Default is ON.
Enabled – Generic data based on that collected during wind tunnel testing is programmed into an internal calibration
table that is used to compensate for turbulence created around the spars and transducers.
Disabled – Turns off the calibration table data, there is no adjustment to take into account of turbulence around the spars
and transducers and returns the unit to raw uncorrected operation.
Gx to Gxxxx – Controls averaging setting (Note, does not apply to Analogue Inputs)
0
Averaging disabled
1-3600
Averaging enabled (time in seconds)
The unit can be used to average readings over long periods of time (up to 3600 seconds). When the G command is used
the unit is automatically switched to P1 mode (1Hz) and will output at the set number of seconds, it will not output any
data in between readings so for long averaging periods the unit will appear to be inactive. If G is set to zero then
averaging settings will be disabled.
While averaging is active the Px command is disabled.
Signal
x
Space
y
Retries Off
1
0
Retries On
1
1
Instantaneous Sampling Off
2
0
Instantaneous Sampling On
2
1
Calibration Off
3
0
Calibration On
3
1