Installation
MetPak Base Station User Manual
Last Change
©
Gill Instruments Ltd
22 April, 2013
Page 15
USB Lead Connection (COMMS set for RS232)
Connect MetPak Base Station to a PC USB port using the Gill Configuration Cable 1723-10-051.
J5 Pin
1
2
3
4
5
6
7
8
Colour
None
None
Yellow
Orange
None
None
Black
Red
Figure 4 Connection of Gill USB Cable (Part No. 1723-10-051)
This USB cable can supply the necessary minimum 5v power for configuring a
Base Station hub box and with a WindSonic/WindSonic M unit connected to the 10
Metre cable.
This USB cable cannot be used to power the Base Station with a WindObserver 70
or WindMaster/WindMaster Pro connected to the 10 Metre cable.
NOTE:
- If upon connecting the USB lead to a PC the driver is not found by the PC, then
the appropriate driver to match the PC Windows version may be downloaded from:-
http://www.ftdichip.com/Drivers/VCP.htm
NOTE.
Ensure that MetPak Base Station 10 Metre cable is terminated before
connecting the USB cable as sensor power is connected to the 10 Metre
cable pins.
Digital Output RS232 Connections (COMMS set for RS232)
This is the MetPak Base Station default communication configuration setting.
J5 Hub Box
8 Way Push fit Block
PC 9 Way D Type
Connector
Signal names
Terminal
Nos.
3 Pair Cable
Pin
No’s
Signal
Names
RS232 TX
RS232 RX
3
4
2
3
RXD
TXD
Signal Ground
2
5
Sig Gnd
Chasis Ground
Terminal
Post
Screen wires
NA
Chassis
Ground
Supply –ve
ve
7
8
–
+
DC Power
supply
NOTE. Ensure that MetPak Base Station 10 Metre cable is terminated before connecting dc
power as sensor power is connected to the 10 Metre cable
Figure 5
RS232 connections
Connect to
Hub box J5