Operation
MetPak Base Station User Manual
©
Gill Instruments Ltd
Last Change
Page 48
22 April, 2013
3.4.4
Configuring MetPak Base Station for SDI-12
Note:
SDI-12 is not supported with a WindMaster or WindMaster Pro connected to
the Base Station.
To configure a MetPak Base Station unit for SDI-12:
1.
Connect a MetPak Base Station to a PC as described in Para 2.2.4.
2.
Open MetSet as described on Page 39.
3.
Click on the Edit Comms page and select SDI-12 from the drop down menu.
4.
Click on Save Settings to Device on the Online Mode Menu.
5.
Shut down the MetPak Base Station unit.
6.
Change the hardware connections for SDI-12.
7.
Power up the MetPak Base Station unit.
3.4.5
SDI-12 Commands
SDI-12 Command Summary
?!
Unit Address (default is 0).
aAb!
Change unit address (a = 0).
aM!
Wind Direction/Speed (direction in Degrees, speed in Metres/Sec),
Temperature (degrees C), Pressure (Hecto Pascals/millibars) and
Humidity (%).
aM1!
Wind Direction/Speed (direction in Degrees, speed in Metres/Sec ),
Temperature (degrees C), Pressure (Hecto Pascals/millibars) and
Dewpoint (degrees C).
aM2!
Wind Direction/Speed (direction in Degrees, speed in Metres/Sec )
and a 2 digit Wind only Status Codes.
aM3!
Temperature (degrees C), Pressure (Hecto Pascals/millibars),
Dewpoint (Degrees C) and full MetPak Base Station Status Codes.
aM4!
Pressure (Hecto Pascals/millibars).
aM5!
PRT, Analogue Input 1, Analogue input 2 and Digital Input 1.
aM6!
PRT.
aM7!
Analogue Input 1 and Analogue Input 2.
aM8!
Digital Input 1.
aM9!
Power Supply Voltage (Volts).
aD0!
Request a line of the above data.