Troubleshooting
MetPak Base Station User Manual
©
Gill Instruments Ltd
Last Change
Page 56
22 April, 2013
Table 12 Status codes as output with the MetPak Base Station data string
Code
Status
Condition
00
OK
No fault conditions detected in
measurement period.
01
Wind Sensor Axis 1 failed
U Axis blocked or faulty.
02
Wind Sensor Axis 2 failed
V Axis blocked or faulty.
03
Wind Sensor Axis 3 failed
W Axis blocked or Faulty
(WindMaster/Pro only).
04
Wind Sensor Axis 1 and 2 failed
U and V axis blocked or faulty.
05
Wind Sensor Axis 1 and 3 failed
U and W axis blocked or faulty
(WindMaster/Pro only).
06
Wind Sensor Axis 2 and 3 failed
V and W axis blocked or faulty
(WindMaster/Pro only).
07
Wind Sensor Axis 1,2 and 3 failed
U, V and W axis blocked or faulty
(WindMaster/Pro only).
08
Wind Sensor NVM error
Non Volatile Memory checksum failed,
data could be uncalibrated.
09
Wind Sensor ROM error
Read Only Memory checksum failed, data
could be uncalibrated.
0B
Wind Sensor reading failed.
Wind Sensor faulty.
10
Hygroclip error
Hygroclip faulty.
20
Dewpoint error
Hub Pec faulty.
40
Humidity error
Hygroclip faulty.
66
Wind Sensor Power
Check Wind Sensor is powered
67
Wind Sensor RS422
Communications
Check RS422 communication wiring
80
Pressure Sensor Warning
Pressure sensor reading not available/unit
faulty.
XX
More than one fault
Fault condition that can be the sum of
0B,10, 20 or 40. e.g. 1B would mean that
both the Wind Sensor (0B) and Hygroclip
(10) were faulty.