The MetPak RG unit stops reporting sensor measurements and reporting (if in
continuous mode) and responds with the following message:
SETUP MODE
The unit is now ready to receive commands.
Returning to measurement mode
To exit from the setup mode and return to the reporting mode:
Type
Q
and press
Enter
.
The MetPak RG unit returns to measurement mode and begins reporting (if set up
for continuous mode).
MetPak RG commands
The following table lists the full set of commands for checking MetPak RG.
Table 10
MetPak RG Commands
MetPak RG
Polling for Data
The MetPak unit must be set for Message Mode Poll using MetSet software
(Comms Page).
A line of data will be output upon receipt of a ? followed by the Node ID (default Q).
It is recommended not to use letters A to F as these characters can appear in the checksum
in the data string.
Notes:-
In Polled mode the unit MUST be allowed 9.5 seconds after switch on to be
operational before the ? and Node letter are used to enable polling or it will
be missed.
When in polled mode, when data is requested allow 570 milli-seconds for the
unit to respond with a line of data.
If the MetPak RG detects a pressure/temperature/humidity sensor fault then
allow 4 seconds for an error data output.
Polled mode can only be used for point to point communications; it is not
possible to network units on to a common bus (except in SDI-12 mode).
Command
Parameters
Description and examples
*
*
Used to place the unit into a SETUP mode so that User Commands may be
sent to review unit settings.
*A-Z
*A | *Z
If the unit is in polled mode this command is used to place the unit into
SETUP MODE so that User Commands may be sent to change unit
settings. Where Q is the default node letter but may be set between A to Z.
It is however recommended not to use letters A to F as these characters
can appear in the checksum in the data string.
?A-?Z
?A to ?Z
?A – ?Z Retrieves a line of data from a MetPak RG set for Polled mode
with a Node identifier set to A to Z.
EXIT
Used to change the unit from Configuration mode to Measurement mode
(alternatively type Q or QUIT)
Q or QUIT
Used to change the unit from Configuration mode to Measurement mode
(see also EXIT)
SERIAL
Used to report the MetPak RG serial number. e.g. 11270001
SWVER
Returns MetPak RG Firmware version e.g. SWVER = 2436 V2.01