Gill Instruments Limited
_____________________________________________________________________________________________________________
________________________________________________________________________________________________
MetStream 100
Page 4
Issue 2
Doc. No. 1955-PS-001
March 2016
3.
SPECIFICATION
MetStream 100 - Parts 1955-100-0X-XX
Input Power Requirement (MetStream only)
Voltage
9-30V DC
Current
145mA max@12V DC (2W) MetStream only
500mA max @ 12V DC, (6W max) including external sensor
Power over Ethernet
Built-in receiver (IEEE 802.3af-2003)
Output Power
Voltage
Regulated 12V DC ± 10%
Current
Up to 300mA @ 12V DC
Digital Input
Communication
ASCII, RS232,RS422/485 4 wire, full duplex
Baud rates
1200, 2400, 4800, 9600, 19200, 38400, 57600
Formats
8 data, odd, even or no parity
Data Input Rate
Up to
32
Hz
Digital Output
Format
8P8C
Communication
Ethernet 10/100 BASE-T speeds/format
Internal Storage
Micro SD Card
8GB card for logged data.
External Data Storage
USB 2.0 Port
Expansion port supplied for external USB Data Storage, max of
32GB supported.
Start Up Time
Data Ready
Less than 90 seconds (MetStream box) from power up.
Dimensions
Size
112mm long, 111mm wide, 34mm deep.
Weight
0.27kg.
Materials
External Construction
Aluminium and Plastic
Environmental
Moisture protection
IP40
IP65 with optional Gill outdoor enclosure
Operating temperature
-20°C to +70°C
Storage temperature
-40°C to +80°C
Humidity
<5% to 95% RH
EMC
BS EN 55024:2010
BS EN 55022:2010
FCC
CFR47 Part 15B
Standards
Traceable to UK national standards