MetStream 100
Page 8
Issue 2
Doc. No. 1955-PS-001
March 2016
4.4.4
Mounting
The MetStream 100 is designed for indoor use.
The MetStream 100 can be positioned laid flat down or on its side.
For outdoor use the unit should be enclosed in a box that provides the necessary
sealing that will meet the outdoor climatic conditions. Gill instruments can provide
an outdoor box if required part number 1955-PK-052.
MetStream Connections.
4.5
4.5.1
MetStream Front Panel Connectors
4.5.2
Front Panel Status LED’s
The Yellow LED illuminated indicates DC Power is applied to the MetStream Box.
The Green LED illuminated indicates unit connecting/connected to a user unit.
A flashing Green LED means that MetStream is communicating with the connected
instrument.
When power is applied to the MetStream with valid equipment wired to the
RS232/422/485 connector the Green LED will be:-
Off initially for approximately 10 seconds.
On for approximately 20 seconds.
Flash whilst correct communication (Comms Mode and Baud Rate) with
the connected device is being established.
On continuously when the connected instrument is recognised and data is
ready to be read.
However allow up to 90 seconds for a valid connection to be established depending on
equipment connected.
4.5.3
Front Panel USB 2.0 Port
Suitable for a USB data stick to copy files from MetStream.
Maximum 32GB device supported.
Hot Pluggable.
Connection
Status LED
USB Port to
copy data
DC Power LED