22
Always use the nebulizer facing upwards in order to prevent any substances and / or the
medicine from leaking out of the nebulizer during normal use.
In the event of overfilling, empty the ampoule, clean it and repeat the operation. After having
poured the medicine, screw the top again to the bottom and repeat the operations as speci-
fied in section “instructions for use”.
WARNING: The power cable plug is the separation component from the power network; even
though the device has the power on / power off switch, the power plug shall be kept accessi-
ble once the appliance is in use in order to allow an additional method of disconnection from
the power network.
NEVER INHALE IN HORIZONTAL POSITION.
DO NOT TILT THE NEBULIZER OVER 60°.
ELECTROMAGNETIC INTERFERENCE RISKS AND POTEN-
TIAL CORRECTIVE MEASURES
This section contains information on the device compliance with EN 60601-1-2 (2015). Cor-
sia Nebulizer is a medical device which is particularly useful for intensive hospital and clinic
use.
CISPR classification: group 1, class B. Corsia Nebulizer is an electro-medical device which
requires particular precautions regarding electromagnetic compatibility and which must be
installed and commissioned according to the electro-magnetic compatibility information sup-
plied.
Avoid using this device close to or overlapped on other appliances because it could not
work properly. If such use is necessary and inevitable, special precautions shall be adopted
so that the electromedical device works properly in its standard configuration (for instance, by
steadily and visually checking the absence of anomalies or malfunctions).
The use of accessories, transducers and cables other those supplied by the manu-
facturer of the appliance may cause an increase in the electromagnetic emissions and/or a
reduction in the electromagnetic immunity of the device, thus causing a malfunction.
Portable and mobile radio communication devices (mobile phones, transceivers, in-
cluding peripheral devices like antennas cables and external antennas, etc.) may affect the
medical device and should not be used close to (at more than 30cm from any part of the
device, including cables), adjacent to or overlapped on the medical device. If such use is
necessary and inevitable, special precautions shall be adopted so that the electromedical
device works properly in its standard configuration (for instance, by steadily and visually
checking the absence of anomalies or malfunctions).
ENGLISH