24
Guide and declaration of the manufacturer – Electromagnetic Immunity
The Corsia aerosol can be used in the following electromagnetic environment. The Cus-
tomer and/or the user of the Corsia aerosol shall make sure that the appliance is used in
such environment.
Immunity
test
Level set out by
EN 60601-1-2
Level of
conformity
Electromagnetic Environment -
Guide
Conducted
immunities
EN 61000-4-6
Radiated
immunity
EN 61000-4-3
3Vrms 150kHz
to 80MHz
(for non-life-
supporting
appliances)
10 V/m 80MHz
to 2.7GHz
(for non-life-equip-
ment appliances)
V
1
=
3 V rms
E
1
=
10 V / m
Portable and mobile RF communication
devices should be used no closer than
the separation distance from any part of
the Corsia device, including cables, cal-
culated from the equation applicable to
the frequency of the transmitter.
Recommended separation distanc-
es
d=
[
3.5
/
V
1
]
√P
d=
[
12
/
E
1
]
√P
80MHz to 800MHz
d=
[
23
/
E
1
]
√P
800MHz to 2,7GHz
P
is the maximum rated output power
of the transmitter in Watt (W) according
to the manufacturer of the transmitter
and
d
is the recommended separation
distance calculated in meters (m).
The field strengths from fixed RF trans-
mitters, as established in an electro-
magnetic survey of the site
a
, could be
less than the level of conformity of each
frequency range
b
.
It is possible to check the
interference close to the
appliances labelled with
the following symbol:
Note 1: At 80 MHz and 800 MHz the highest frequency range applies.
Note 2: These guidelines could not apply to all conditions. The electromagnetic propaga-
tion is affected by the absorption and reflection of premises, objects and individuals.
ENGLISH
Network frequency
magnetic field
EN 61000-4-8
30 A/m
The appliance
does not alter
its status
The power supply should be
that of a typical commercial
premise or hospital.
Note:
U
T
is the value of the supply voltage