facturer of the appliance may cause an increase in the electromagnetic emissions and/or a 

reduction in the electromagnetic immunity of the device, thus causing a malfunction.

 Portable and mobile radio communication devices (mobile phones, transceivers, in-

cluding peripheral devices like antennas cables and external antennas, etc.) may affect the 

medical device and should not be used close to (at more than 30cm from any part of the 

device, including cables), adjacent to or overlapped on the medical device. If such use is 

necessary and inevitable, special precautions shall be adopted so that the electromedical 

device works properly in its standard configuration (for instance, by steadily and visually 

checking the absence of anomalies or malfunctions).

The tables below provide information on the EMC characteristics (Electromagnetic Compat-

ibility) of this electromedical appliance.

ENGLISH

Guide and declaration of the manufacturer – Electromagnetic Emissions 

The Mistral aerosol can be used in the following electromagnetic environment. The Cus-

tomer and/or the user of the Mistral aerosol shall make sure that the appliance is used in 

such environment.

Emission test

Conformity

Guide to the electromagnetic 

environment

Radiated emissions / 

Conductions CISPR11 

Group 1

The Mistral aerosol uses RF energy only for 

Internal function. Therefore, its RF emissions 

are very low and do not cause any interfer-

ence with other nearby electronic applianc-

es.

Radiated emissions / 

Conductions CISPR11 

Class [B]

The Mistral aerosol is designed for use in 

any environment, including households and 

those directly connected to the public power 

distribution grid which supplies power to en-

vironments intended for domestic use.

Harmonic currents EN 

61000-3-2

Class [A]

Voltage fluctuations / 

flicker EN 61000-3-3

Compliant

Guide and declaration of the manufacturer – Electromagnetic Immunity

The Mistral aerosol can be used in the following electromagnetic environment. The Cus-

tomer and/or the user of the Mistral aerosol shall make sure that the appliance is used in 

such environment.

Immunity test

Standard of proof

Level of 

conformity

Guide to the electromag-

netic environment

Electrostatic 

discharge (ESD) 

EN 61000-4-2

± 8kV contact

± 15kV air

The appliance 

does not alter 

its status 

Floors should be in wood, 

cement or ceramics. If floors 

are covered by synthetic 

material, the relative humid-

ity should be at least 30%.

Fast transient / burst

EN 61000-4-4

± 2kV power supply

± 1kV signal cables

The appliance 

does not alter 

its status

The power supply should be 

that of a typical commercial 

premise or hospital.

Summary of Contents for 28102

Page 1: ...the product AVIS Les op rateurs doivent lire et bien comprendre ce manuel avant d utiliser le produit ACHTUNG Die Bediener m ssen vorher dieses Handbuch gelesen und verstanden haben bevor sie das Pro...

Page 2: ...nected 3 Do not leave the appliance plugged unnecessarily disconnect the plug from the power network if not used 4 Comply with the safety standards for electrical appliances In particular Use only ori...

Page 3: ...people with disabilities 10 Do not leave the device unattended in areas accessible to minors and or people with limited mental capacity as they may be strangled with the air hose 11 The patient can c...

Page 4: ...to any of its part due to accident misuse and or abuse Any unauthorized intervention on the device even the slightest one will imme diately invalidate the warranty and does not ensure the compliance w...

Page 5: ...detergents Do not use excessively wet cloths as the contact of liquids with the electri cal parts of the appliance may cause malfunctions or may be hazardous While cleaning the device make sure that n...

Page 6: ...ntact the reseller or Gima s technical service BEFORE CARRYING OUT ANY CHECKS IN THE EVENT OF ANOMALIES OR MALFUNCTIONS CONTACT GIMA S TECHNICAL SERVICE THE MANUFACTURER OFFERS NO GUARANTEE FOR THE AP...

Page 7: ...WASHING Where diseases with risk of infection and microbial contamination are present the end user shall carry out the washing operations properly The washing procedure can be performed only if the co...

Page 8: ...or if the device is no longer efficient If the filter is too dirty it shall be replaced AIR FILTER REPLACEMENT The air filter shall be replaced every 25 hours of operation or when it is too dirty For...

Page 9: ...e nebulizer facing upwards in order to prevent any substances and or the medicine from leaking out of the nebulizer during normal use In the event of overfilling empty the ampoule clean it and repeat...

Page 10: ...e electromagnetic environment Radiated emissions Conductions CISPR11 Group 1 The Mistral aerosol uses RF energy only for Internal function Therefore its RF emissions are very low and do not cause any...

Page 11: ...ower of the transmitter in Watt W according to the manufacturer of the transmitter and d is the recommended separation distance calculated in meters m The field strengths from fixed RF trans mitters a...

Page 12: ...eparation distances between portable and mobile radio communi cation appliances and the monitor The Mistral aerosol is designed to operate in an electromagnetic environment in which the RF radiated in...

Page 13: ...expenses and packaging not included All components subject to wear are not included in the warranty The repair or replacement performed during the warranty period shall not extend the warranty The wa...

Reviews: