127
ENGLISH
VO
2
calc (oxygen consumption), C(a- v)O
2
(oxygen content difference between artery and vein), O
2
ER (oxygen extraction
rate), DO
2
(oxygen delivery), PAO
2
(oxygen pressure in alveolar ), AaDO
2
(oxygen pressure difference between alveolar
and artery), CcO
2
(pulmonary capillary oxygen content), Qs/Qt (intra-pulmonary shunt rate), C.O.calc (calculated cardiac
output), PaO
2
/ FiO
2
(oxygenation index), AaO
2
/ PaO
2
(Ratio of alveolar-arterial oxygen partial pressure difference to
oxygen partial pressure), DO
2
l (oxygen delivery index) and VO
2
l (oxygen consumption index).
Refer to the following table for the detailed definition, unit and adjustable range of the output items.
Parameters
Definition
Calculation formula
Unit
Range
BSA
Body surface area
W
0.425
*H
0.725
*0.007184
m
2
——
VO
2
calc
Oxygen consumption
C ( a- v ) O
2
*13.4*Hb * C.O.
ml/min
——
C(a- v)O
2
Oxygen content difference
between artery and vein
CaO
2
– CvO
2
ml/L
42~59
O
2
ER
Oxygen extraction rate
VO
2
(CaO2-CvO2)*100
%
24~28
DO
2
Oxygen delivery
CaO
2
* C.O.
ml/min
950~1150
PAO
2
Oxygen pressure in alveolar (FiO
2
)*(ATMP - 47) – (PaCO
2
/ RQ)
mmHg
——
AaDO
2
Oxygen pressure difference
between alveolar and
artery
(FiO
2
)*(ATMP - 47) – (PaCO
2
/ RQ)
-
PaO
2
mmHg
10~15
CcO
2
Pulmonary capillary oxygen
content
(Hb*1.34)+(((FiO
2
)*(ATMP - 47) – (PaCO
2
/ RQ))* 0.0031)
ml/L
——
Qs /Qt
Intra-pulmonary shunt rate
{[(Hb*1.34)+(( (FiO
2
)*(ATMP-47)–
(PaCO
2
/RQ))* 0.0031)]
-
CaO
2
) / {
[(Hb*1.34)+(( (FiO
2
)*(ATMP-47)–(PaCO
2
/
RQ))* 0.0031)] – CvO
2
}*100
%
3.0~5.0
C.O.calc
Calculated cardiac output
VO
2
/ (CaO
2
-CvO
2
)
L/min
0.1~20.0
PaO
2
/ FiO
2
Oxygenation index
PaO
2
/ (FiO
2
)
mmHg
——
AaO
2
/ PaO
2
Ratio of alveolar-arterial
oxygen partial pressure
difference to oxygen partial
pressure
[(FiO
2
)*(ATMP - 47) – (PaCO
2
/ RQ)
-
PaO
2
] / PaO
2
——
——
DO
2
l
Oxygen delivery index
(CaO
2
* C.O.)/
BSA
ml/min/m
2
——
VO
2
l
Oxygen consumption index (CaO
2
– CvO
2
)*C.O. /BSA
ml/min/m
2
——
Operations:
Entering the value for each input item, and click “Calculate”, and the monitor will calculate the parameter according to
the calculate formula, the result will be displayed on the screen. then click “Range” to view the range for each parameter.
Notes:
1) The result for each parameter will be shown as “---” before click “Calculate” button.
2) The calculation result will be shown in yellow if the result is beyond the reference range.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...