94
ENGLISH
13.3 CO
2
Sensor Connection
13.3.1 Sidestream CO
2
Sensor Connection
Demonstration for Sidestream CO
2
Sensor Connection
1. Take out the CO
2
Sensor and insert the CO
2
Sensor Cable into the connector labeled “CO
2
” on the connector panel of
the monitor;
2. The sample cell of the sampling cannula must be inserted into the sample cell receptacle of the CO
2
Sensor. A “click”
will be heard when the sample cell is properly inserted, then connect to the airway tube. After finishing sensor
connection, and make sure that the air input end is exposed to room air and away from all sources of CO
2
, including
the ventilator, the patient’s breath and your own. Next, turn on the CO
2
switch at CO
2
Setup Screen and then wait 2
minutes for the sensor warm-up.
3. Default Tubing Configuration
Sampling adapter
Dehumidification
tubing
Sampling tubing
Filter
Sampling cell
Adapter and Sampling tube (Single patient use)
Extending airway tube for connecting to sampling tube (Single patient use)
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...