Operator
’
s Manual
MO080-IT
12/09/2017
Rev.3
Page 9 of 20
in the terminal marked with the ground symbol by means of the nut.
Product power protection is ensured by an input fuse (L, N) of the T1AH 250V 5X20 type, which are already
connected in the switchboard.
2.4
Handpiece fitting
To fit the handpiece, turn it clockwise inside the threaded hole provided until it is up against the headpiece and
rotation remains blocked.
2.5
First switch-on
At this point, the Product can be switched on to make sure it works properly. Follow the instructions below:
1-
Switch on the power to the Product by means of the master switch.
2-
Press the I/O keyboard located on the front part of the reflector.
3-
Make sure all LEDs and functions are working properly.
2.6
Check the result of Product installation and testing before use
The following instructions are to be deemed mandatory during the installation inspection phase, as they prove that the
various jobs referred to have been correctly done. Hence each single step must be ticked.
1.
Make sure the ceiling is suitable for Product installation.
2.
With the aid of a spirit level, make sure the bar tube is perfectly perpendicular
3.
Make sure the bolt is tight on the stop collar
4.
Make sure the hole has been drilled correctly and that the safety dowel is fitted on the bar tube
5.
Make sure the 4 x M4 screws which sustain the horizontal arm are tight
6.
Check the Product earth connection and make sure the clamps are well tightened
7.
Make sure movement mechanisms are working properly.
8.
After switch-on the Product must emit light from the headpiece
Installer’s stamp and signature:
3
Importance of personal safety
3.1
Intended use
The Product has been designed to light up the area of the patient undergoing observation and diagnosis and is
intended for use in doctors’ surgeries.
The Product correctly lights up the operating field from a minimum distance of 40 cm and a maximum distance of
about 70 cm, from the point of light emission.
The Product, in conformity with the IEC 60601-2-41 standard, is defined as a lamp for diagnostics:
-
A lamp for diagnostics is a lamp used to locally light up the body of a patient, in order to make diagnosis or
treatment easier. These can be interrupted without any danger for the patient in case of the light going off. (The
Product is not intended for use in operating theatres).
3.2
Environmental conditions
-
The Product is not suitable for use in explosion-risk areas.
-
The Product is not suitable for use wherever there are inflammable mixes of anaesthetics with air, oxygen or NO
2
(laughing gas).
-
The Product is not suitable for use in environments rich in oxygen and use is not intended in the presence of
inflammable agents.
-
During operation, the ambient temperature must be between 10°C and 40°C.
-
Relative humidity must be between 30% and 75%.
-
Atmospheric pressure must be between 700 and 1060hPa.
Summary of Contents for PRIMALED SO
Page 15: ...Operator s Manual MO080 IT 12 09 2017 Rev 3 Page 15 of 20 8 Wiring diagram ...
Page 21: ...Drawing code Rev Data 052 1 2 3 ...
Page 22: ...Drawing code Rev Data 053 1 2 3 ...
Page 23: ...Drawing code Rev Data 054 2 1 3 4 ...
Page 25: ...Drawing code Rev Data 056 2 1 3 4 5 ...
Page 26: ...Drawing code Rev Data 057 3 1 2 ...
Page 27: ...Drawing code Rev Data 058 2 1 3 4 ...
Page 28: ...Drawing code Rev Data 1 2 7 3 4 5 6 153 ...
Page 29: ...Drawing code Rev Data 190 1 2 3 4 5 6 8 7 ...
Page 30: ...Drawing code Rev Data 191 1 2 4 5 3 6 7 ...
Page 31: ......
Page 32: ......