37
EN
EN
USER MANUAL
BAR
P2 CAL
Normal calibrate
After enter cal mode, the display will show UnLoAd. Remove all weight from the pan and wait for
stable.
After stable, balance will show LoAd, put calibrate weight on the pan, here is the useable weight list.
300g:
30/40/50/60/70/80/90/100/110/120/130/140/150/160/170/180/190/200/210/220/230/240/250/
260/270/280/290/300
600g:
60/80/100/120/140/160/180/200/220/240/260/280/300/320/340/360/380/400/420/440/460/480
/500/520/540/560/580/600
1500g:
150/200/250/300/350/400/450/500/550/600/650//700/750/800/850/900/950/1000/1050/1100/1 150/12
00/1250/1300/1350/1400/1450/1500
3000g:
300/400/500/600/700/800/900/1000/1100/1200/1300/1400/1500/1600/1700/1800/1900/2000/21 00/22
00/2300/2400/2500/2600/2700/2800/2900/3000
after stable, dosplay will show pass, calbrate complete.
No need press any jey during calibrate.
P3 Cnt
This display will show xxxxxx for indicating the internal counts, you can press
key to escape.
P4 A2n
This option is used to select the auto zero maintain
Options: OFF, 0.5d, 1d, 2d, 4d, Use
key to shift, press
key to sure, press
key to scape.
P5 GrA
This is used to adjust gravity, when move to another place with big latitude change, you can use this
function to adjust.
After press
key to enter this function, you can use
,
,
key to key in the latitude rate
(Use
and
key move active digits, use
key incease value), after input, press
key to sure.
Latitude rate= local latitude/the latitude of the calibrate place.
For example, local latitude is 9.85, the latitude of the calibrate place is 9.75, by this way, the read-ing
data will be bigger than standard, you need input 1.01025, then reading will be correct. Note: after you
do calibrate, this rate will reset to 1.00000.
P6 CAP
Set capacity, press
key to select: 300g/600g/1500g/3000g/6000g, for high resolution model, you
can select 150g/200g/1500g/2000g, after you select, press
to confirm.
P7 SPD
Converter speed L / m / H (low / medium / high)
P8 S-t
Multi-tare ON / OFF
P9 ret
Ok