USER MANUAL
BAR
9. APPENDIX
9.1. ERROR CODES
During the initial power-on testing or during operation it is possible the scale may show an error message. The meaning of the
error messages is described below.
If an error message is shown repeat the procedure that caused the message, turning the balance on, calibration or other functions.
If the error message still is shown then contact your dealer for further support.
ERROR CODE
DESCRIPTION
POSSIBLES CAUSES
Err 3
Calibrate weight error
Use incorrect calibrate weight
Err 4
Initial Zero is greater than allowed (4% of
maximum capacity) when power is turned
on or when the
key is pressed.
Weight on the pan when turning the scale
on.
Excessive weight on the pan when zeroing
the scale.
Improper calibration of the scale. Damaged
load cell.
Damaged electronics.
Err 6
A/D count is not correct when turning the
scale on
Platform not installed.
Load cell damaged.
Electronics damaged.
Err 7
Percentage setting error
0.01% weight must>0.5d
Err 8
External calibration weight error
More than 4% error as linear calibration
9.2 WEIGHING UNIT LISTING
Nº
UNIT SYMBOL
NAME OF THE UNIT
CONVERSION
1
g
gram
2
g
gram (high resolution)
3
ct
carat net
= 0,2 g
4
lb
pound
= 453,59237 g
5
oz
ounce
= 28, 349523000 g
6
d
drachm
= 1,7718450 g
7
ozt
ounce troy
= 31,1034768 g
8
dwt
scruples
= 1,55517400 g
9
mom
momme
= 3,749996 g
10
tL.H
tael twn
= 37,49996 g
11
tl.c
tael chn
= 37,799375 g
12
tl.t
troy tael
= 37,4290018 g
13
t
tola
= 11,6638039 g
14
bt
bangladeshi tola
1bt= 11,6638039 g
1bt= 16Ana= 96Roti
15
n
newton
1n= 102,04 g
10. TROUBLE SELF CHECKING
PROBLEM
POSIBLE CAUSE
COMMON SOLUTIONS
Display is blank, No turn on test
Mains power is turned off.
Power supply not plugged in.
Power supply faulty.
Internal battery not charged.
Display turned off.
Check power is going to the scale and
switch is on.
Verify the voltage going to the scale
matches the power supply labels on
the power module or scale.