Display blank after turn on
test, Error message, Display is
locked
Pan not installed.
Unstable weight.
Load cell damaged.
Mechanics damaged.
Check the pans are installed correctly.
Try turning the scale on again.
OL or ---- appears on display
Maximum capacity exceeded. Load cell
or mechanics damaged. Power supply
faulty.
Check the pans are installed correctly.
Try turning the scale on again.
---- or NULL displayed
Weight on scale below permissible
li-mit.
Pan has been removed.
Pan support not seated properly. Power
supply faulty.
Load cell or mechanics damaged.
Check the pans are installed correctly
Try pressing
.
Try turning the scale on again.
Display is unstable
Drafts or air currents.
Obstruction under pan.
Sample is moving (animal weighing).
Vibrations through table. Temperature
changed dramatically. Power supply
faulty.
Verify the scale is in a acceptable loca-
tion and on a good table.
Verify the power supply is correct for
the scale.
Weight value incorrect
Calibration error, Recalibrate. Linearity
error, set Linearity.
Unit calibrated with inaccurate weight.
Balance not level.
Obstruction between sample and
co-ver.
Wrong unit of weight displayed.
Calibrate again, paying special atten-
tion to the mass used, the stability of the
scale, and the weighing units re-quired.
If linearity can be set by the user it
will be described in the user manual.
Check pan is installed correctly. Verify
installation is acceptable.
Cannot use Full Capacity
Over load Stops hitting pan support or
hitting bottom of load cell.
Shipping screw not removed if
appli-cable.
Electronic problem on A/D. Parameters
set incorrectly.
Load cell or mechanics damaged.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-tion.
Check the weighing units used.
Not linear
Overload stops hitting too soon.
Load cell or mechanis damaged.
A/D damaged.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-tion.
If linearity can be set by the user it will
be described in the user manual.
Off center
loading error
Adjust mechanics.
Overload stops not correct.
Load cell damaged.
Look for obstruction under pan, ship-
ping screws, and verify pan installa-tion.
Battery will not charge
Mains voltage not present or too low.
Charging circuit failure.
Battery failure.
Verify the batteries are rechargeable
types.
Check power supply voltage is correct.
11. WARRANTY
This scale is warranted against defects of manufacturing and materials for a period of 1 year, from the delivery date.
During this period, GIROPÈS SL will take charge of repairing the scale
This warranty does not cover defects or damaged caused by misuse, overloading or improper installation contrary to the recom-
mendations described in this manual.
This warranty does not cover shipping costs for the reparation of the balance.