Operating Instructions
5.1 Connecting the LCR-600
Connect the power cable to your meter. Now connect the BNC
Plug to Clip Lead Wire to the BNC terminals on the LCR-600. Make
sure that you connect according to the color bands.
5.2 Powering On
LCR-600 has two power switches: Line Power Switch [25] on
the rear panel for the transformer and then the Power Button [3] on
the front panel for the operational system.
Switch “ON” the Line Power Switch [25] on the rear panel then
press the Power Button [3] on the front panel to light the LCD.
5.3 Open Circuit / Short Circuit Compensation
LCR-600 provides open circuit and short circuit compensation
so that you can measure high resistance and low resistance more
accurately. You must first calibrate these functions. Start by making
sure the clip ends of the leads are not touching each other. Set the
display at “OPEN” by pressing the CAL [6] key for 2 seconds. The
main LCD will display [Open]. Then press CAL [6] again to start the
open circuit calibration. The calibration will need about 30 seconds,
after which the LCD will display [PASS]. The LCR-600 has now
automatically finished the open circuit calibration.
Now for the short circuit calibration, connect the two clip ends
of the lead together to create a short circuit. Press the CAL [6] key
again for 2 seconds. The main LCD will display [Srt]. Now press the
CAL [6] key again to start the short circuit calibration. As with the