23
9
Open / Short
Calibration
After Step.8, the LCD will display [OPEN]. Keep 2
input tip at “OPEN” condition and press CAL key.
The LCD will flash 30sec and display [PASS], then
short 2 input tip, the LCE will display [SRT]. Press
CAL key, after 30sec. flash, the LCD will display
[PASS], then go to step.10. If the LCD display
[FAIL], repeat step.9 again.
10
1KHz
10M
Ω
10.000M
Ω
The same operation as
step.1.
11
1KHz
1M
Ω
1.0000M
Ω
The same operation as
step.1.After step.11, the
LCR-600 will changed to
10KHz automatically. Go
to step.12.
12
10KHz
1M
Ω
1.0000M
Ω
The same operation as
step.1.
13
10KHz
100K
Ω
100.00K
Ω
The same operation as
step.1. After step.13, the
LCD will display [100KHz]
Go to step.14
14
100KHz
100K
Ω
100.00K
Ω
On this step, the
resistor should be 100K
Ω
/ 100KHz standard
resistor. Operate as
step.1
15
100KHz
10K
Ω
10.000K
Ω
Operate as step.1
16
100KHz
10
Ω
10.000
Ω
Operate as step.1.
17
100KHz
1
Ω
1.0000
Ω
Operate as step.1
18
After finished the calibration. The unit will power off automatically.
Then please switch “OFF” the power switch and “OPEN” J11 and
cover the cabin.
19
After switching “ON” the unit again, the unit will return normal
operation mode. Then press CAL key 2 sec to operate open
circuit/short circuit calibration.