244280
www.globalindustrial.com
12
incorrect data input or loading weight), please return back to the last step or use
ESC/ON/OFF
to exit the calibration mode.
6. Parameter Setup Mode
6.1 In normal weighing mode, press and hold
ESC/ON/OFF
and
2
key until scale
displays “SETUP”.
6.2
Then scale displays “A.O.T xx”, it means you have entered parameter setup mode.
6.3 In setup mode, use
0
-
9
key to input numbers, use
C
key to clear the input value,
use
TARE
key to confirm and go to the next parameter, use
ESC/ON/OFF
key to exit the
mode.
Parameter menu:
Paramet
er
meaning
option
Default
setting
A.O.T xx
Auto off time
setting
00~30 minutes
5
BLT. x
Blacklight
0 - Backlight always off
1 - Backlight always on
2 - Auto on when key operation or weight
changes; auto off if no operation in 15s
2
CST. x
LCD
contraction
level selection
0, 1, 2, 3,
4
, 5, 6, 7
4
ACC.xxx
Accumulation
mode
0-MAN, manual accumulation
1-AUT, auto accumulation
MAN
BEEP.xx
Beeper mode
in check
weighing mode
0-beeper always off
2
1-compare result is LO, beeper on
2- compare result is OK, beeper on
3- compare result is HI, beeper on
BPS.xxx
Baud rate
1200bps, displays as “12”
2400bps, displays “24”
4800bps, displays “48”
9600bps,
displays as “96”
19200bps, displays as “192”
96
DFT.xxx
Byte format
1 - 8N1=8 data bits, No parity check bit, 1 stop bit;
2 - 7O1=7 data bits, 1 parity check bit, 1 stop bit;
3 - 7E1=7 data bits, 1 parity check bit, 1 stop bit;
1
CFT. x
Output mode
0-no communication
1-output after reading is stable
2-output after PRINT key is pressed
3-output after reading is stable or after PRINT key is
pressed
2