GLP Santinel User Manual
6
•
Ensure that the product is electrically connected to ground (earth).
•
Disconnect the product from AC mains power before carrying out service or moving
the product and when the product is not in use.
•
Disconnect the product from power if the power plug, power cable or any cover or
other component is damaged, defective or deformed. Do not reapply power until
the product has been repaired and made safe by a technician authorized by GLP.
•
If a fuse blows, replace it with one of the original type and rating only. If new fuses
blow, disconnect the product from power and send it to a technician authorized by
GLP for repair. Do not attempt to bypass a fuse.
Disinfectant safety
•
Use only a disinfectant that is approved by GLP.
•
Do not use the disinfectant on babies under 28 days old or premature babies.
•
The disinfectant is harmful when swallowed. Do not allow any person to drink it. If a
person has drunk hand disinfectant, give him a glass of water or tea to rinse, so that
the disinfectant is diluted. If necessary, give him something sweet to keep the blood
sugar from dropping. In any case, call medical assistance.
•
If you need to transfer disinfectant from one container to another, do so under
aseptic conditions only.
•
For information on risks and possible side effects in connection with the disinfectant,
read the package insert supplied with the disinfectant container. If in any doubt,
consult a doctor or pharmacist.
Installation safety and protection from personal injury
•
Ensure that the product is safely supported and cannot fall. If mounting the product
on a wall or other vertical surface or structure, use fasteners that are suitable for the
surface and strong enough to safely hold the weight of the product with a full
container of disinfectant plus a generous safety margin so that there is no risk of the
product falling.
•
Install the product on a surface, stand or structure only as directed in this user
manual.
•
Check that the hatch over the disinfectant container is securely locked before
allowing the product to be used.
•
If the product becomes damaged, stop using it immediately and disconnect it from
power. Do not attempt to use a product that is obviously damaged.
•
Do not modify the product in any way not described in this user manual.
Summary of Contents for Santinel Cellular
Page 30: ...www glp de GLP Santinel User Manual 30 12 Dimensions...
Page 31: ...Base Access flap...
Page 32: ...Wall Mount Plate...
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......