Troubleshooting 

 

/

 

English

Device in water or other liquids. If this occurs, dry off quickly with a cle-

an cloth.

 Do not store the Device in a freezer.

 Do not use the Device if abnormal heat, odour, discolouration, defor-

mation, or other abnormal condition is detected during use, charging, 

or storage.

/ 3.14

 

Long-term storage

If you decide not to use your GlucoMen Day PENCAP and wish to store it for 

future use:

1.

 

Fully charge the battery.

2.

 

Power off the Device (follow steps in SECTION 3.10).

a.

 Do not mount the Device on a pen after powering off.

b.

 Do not click the sleeve (without pen) to engage the mounting mecha-

nism – this will power on the Device.

3.

 

Store the Device at temperatures between -10 to 60 °C (14 to 

140 °F) in Relative Humidity of 45-85%, at altitudes of < 3000 m 

(< 9840 ft.) and at normal Atmospheric Pressure.

/ 3.15

 

Disposal

At the end of the Device’s service life it should be disposed as electrical 
waste according to local regulations.

/ 3.16

 

General safety advice

 

CAUTION

 

  Only  use  a  certified  5V  USB  charger  from  a  legitimate  supplier  with 

the Device (for example, CE marked etc.) to minimise the risk of electric 

shock and damage to the Device.

 In the unlikely event of deterioration in the functionality of nearby equipment 

in the presence of GlucoMen Day PENCAP, remove the Device from the vicinity.

 To avoid degradation in performance of GlucoMen Day PENCAP, radio 

equipment (such as phones, computers, wireless devices, antennas, an-

tenna  cables  etc.)  should  not  be  used  closer  than  30  cm  (12  inches) 

from the Device. This distance should also be observed for the charging 

cable if it is plugged-in to the Device.

  To  reduce  the  risk  of  interference  from  outside  sources,  avoid  using 

GlucoMen Day PENCAP near strong sources of electromagnetic radiation 

(e.g. CT, MRI, X-ray equipment etc.).

 To ensure proper operation of the Device, avoid using it adjacent to or 

stacked with other electrical equipment.

 

Do not use the Device in oxygen rich environments.

/ 4

 

Troubleshooting

/ 4.1

 

Injection not detected

The main cause for injections not detected is that the optical sensor in 

the Device is not working properly. The following can be helpful in fixing 

the issue:

 Make sure that the Device is designated for the pen model it is mounted 

on (see SECTION 2.5).

 Make sure that the Device is fitted and positioned correctly on the pen 

i.e. the optical sensor is positioned close to the dose-knob after moun-

ting (see SECTION 3.1).

 Press the mounting mechanism together to ensure that the Device is 

properly mounted on the pen. The ‘mounted on pen’ animation appears 

when the Device is properly mounted (see SECTION 3.1).

 Make sure that there is no build-up of dirt etc. on the optical sensor 

(see SECTION 3.12 for cleaning instructions).

 If it is still not working, unmount and re-mount the Device on the pen 

(see SECTION 3.8).

• 

If injections are still not detected contact A. Menarini Diagnostics Cu-

stomer Service (reported on the box).

Summary of Contents for Day PENCAP

Page 1: ...User Guide...

Page 2: ...If you have any questions about this product please consult our website www glucomenday com CAUTION Read all instructions and warnings before use A Menarini Diagnostics only accepts responsibility fo...

Page 3: ...Maximum number of pairings 3 8 Clearing paired devices 3 9 Changing Pen 3 10 Factory reset 3 11 Powering off 3 12 Powering on 3 13 Cleaning and care 3 14 Long term storage 1 1 1 1 1 1 1 2 2 3 3 4 4 5...

Page 4: ...you have to decide when to perform an insulin injection The Device is designed to be used in common domestic environments such as your home office or school 1 4 Intended Users GlucoMen Day PENCAP is i...

Page 5: ...kaging is damaged or opened Degree of enclosure protection Protect Device from moisture Avoid using the Device under direct sunlight Range of temperature upper and lower limit to which the Device can...

Page 6: ...2 3 GlucoMen Day PENCAP terminology The following images describe the key components of Insulcheck Connect 1 Display screen Timer and other information is displayed on it 2 Charging indicator Display...

Page 7: ...odel of pen The abbreviation for the pen name the Device is designated for is embos sed on the sleeve The following table provides a list of pen models supported by GlucoMen Day PENCAP together with t...

Page 8: ...tions 3 1 First time use 1 Peel off the labels a First from underneath the Device and then from the touch area and di splay CAUTION Part of this label prevents acciden tal powering on of the Device du...

Page 9: ...actual injections Its operation is based on sensing the changes in the position of the dose knob when the user interacts with it during injections Playing or fiddling with the dose knob of the pen is...

Page 10: ...ons is stored on the De vice for up to 30 days You can transmit this information to the GlucoLog RapidCalc app via Bluetooth 1 On your app or phone activate Bluetooth and scan for available devices 2...

Page 11: ...t smartphones or apps When you pair with a four th client device your GlucoMen Day PENCAP will automatically drop the first client device you connected with to accommodate the new client You can clear...

Page 12: ...the GlucoMen Day PENCAP memory as well as the pairing data can be deleted by doing a Factory Reset GlucoMen Day PENCAP also reboots as a result of a factory reset 1 Unmount the Device from the pen 2 T...

Page 13: ...vice can be powered on by the following methods Connecting the Device to a charger Mounting the Device on a pen Clicking the sleeve without pen to engage the mounting mechanism 3 13 Cleaning and care...

Page 14: ...Day PENCAP radio equipment such as phones computers wireless devices antennas an tenna cables etc should not be used closer than 30 cm 12 inches from the Device This distance should also be observed...

Page 15: ...5 Unsuccessful pairing There are a number of reasons for an unsuccessful Bluetooth pairing The most common reasons are 1 It may be that the passkey was mistyped into the app or pho ne Repeat the proc...

Page 16: ...warning is displayed for a few seconds before displaying the timer When a charger is connected to the Device and the battery is not fully charged this ani mation is displayed for a few seconds before...

Page 17: ...Zed Ltd NovaUCD Belfield Innovation Park Belfied Dublin 4 Ireland By Scandinavian Healthcare Ltd Taiwan 6 3 Warranty Your GlucoMen Day PENCAP is guaranteed to be free of material and workmanship defec...

Page 18: ...function or are da maged due to obvious tampering misuse alteration neglect unau thorized maintenance or failure to operate the GlucoMen Day PENCAP in accordance with the instructions There is no othe...

Page 19: ...on IC Industry Canada Radio Module IC ID 7922A 4008 The Device meets the requirements for modular transmitter approval as detailed in RSS GEN Operation is subject to the following two condi tions 1 Th...

Page 20: ...www glucomenday com 51971 09 20 EN...

Reviews: