12

Display guide 

 

/

 

English

/ 4.2

 

Blinded sensor

The  optical  sensor  of  the  Device  is 

blinded  when  exposed  directly  to 

intense  light  (e.g.  bright  sunlight) 

and cannot detect the activity of the 

dose-  knob.  The  screen  will  display 

the blinded sensor warning.

If  this  happens,  take  the  device  out 

of the intense light until the warning 

is  no  longer  displayed.  The  Device 

can now be used normally.

/ 4.3

 

Spillage

Should  an  accident  occur  exposing  GlucoMen  Day  PENCAP  to  a  liquid 

(e.g. a cup getting knocked over causing spillage), remove it from the 

pen, dry it by wiping gently with a clean dry cloth, and set the Device 

aside for several hours before resuming its use.

 

CAUTION

 

The Lithium-Ion battery should not be exposed to water or other liquids.

/ 4.4

 

System error

If  the  InsulCheck  insignia  is  di-

splayed  periodically  on  the  screen, 

this means that the Device has a re-

peating system error which it cannot 

recover  from.  If  that  happens,  stop 

using the Device and contact A. Me-

narini Diagnostics Customer Service 

(reported on the box). 

/ 4.5

 

Unsuccessful pairing

There are a number of reasons for an 

unsuccessful Bluetooth

®

 pairing.

The most common reasons are:

1.

 

It may be that the passkey was mistyped into the app or pho-

ne. Repeat the process or follow the instructions on your phone 

or app.

2.

 

Check that your smartphone supports connection to your 

GlucoMen Day PENCAP, and that you are using the GlucoLog 

RapidCalc app.

3.

 

The Bluetooth

®

 functionality is implemented in different ways 

by mobile device manufacturers. There may be a incompatibility 

issue which cannot be resolved.

If you still cannot pair with the desired client device, please contact 

A. Menarini Diagnostics Customer Service (reported on the box).

/ 5

 

Display guide

Graphic

Reference 

GlucoMen  Day  PENCAP  insignia  is  displayed 

on the screen for a few seconds after the De-

vice is powered on. It is also displayed after 

a  factory  reset.  The  insignia  is  generally  fol-

lowed by displaying the Firmware Version on 

the Device for a few seconds.

When the screen is turned on, the timer is di-

splayed  on  it.  The  timer  counts  up  the  time 

since the previous dialling-in of the dose- knob 

(during injection) was detected. The timer di-

splays the time since the previous injection.

Insigna displayed

periodically

Summary of Contents for Day PENCAP

Page 1: ...User Guide...

Page 2: ...If you have any questions about this product please consult our website www glucomenday com CAUTION Read all instructions and warnings before use A Menarini Diagnostics only accepts responsibility fo...

Page 3: ...Maximum number of pairings 3 8 Clearing paired devices 3 9 Changing Pen 3 10 Factory reset 3 11 Powering off 3 12 Powering on 3 13 Cleaning and care 3 14 Long term storage 1 1 1 1 1 1 1 2 2 3 3 4 4 5...

Page 4: ...you have to decide when to perform an insulin injection The Device is designed to be used in common domestic environments such as your home office or school 1 4 Intended Users GlucoMen Day PENCAP is i...

Page 5: ...kaging is damaged or opened Degree of enclosure protection Protect Device from moisture Avoid using the Device under direct sunlight Range of temperature upper and lower limit to which the Device can...

Page 6: ...2 3 GlucoMen Day PENCAP terminology The following images describe the key components of Insulcheck Connect 1 Display screen Timer and other information is displayed on it 2 Charging indicator Display...

Page 7: ...odel of pen The abbreviation for the pen name the Device is designated for is embos sed on the sleeve The following table provides a list of pen models supported by GlucoMen Day PENCAP together with t...

Page 8: ...tions 3 1 First time use 1 Peel off the labels a First from underneath the Device and then from the touch area and di splay CAUTION Part of this label prevents acciden tal powering on of the Device du...

Page 9: ...actual injections Its operation is based on sensing the changes in the position of the dose knob when the user interacts with it during injections Playing or fiddling with the dose knob of the pen is...

Page 10: ...ons is stored on the De vice for up to 30 days You can transmit this information to the GlucoLog RapidCalc app via Bluetooth 1 On your app or phone activate Bluetooth and scan for available devices 2...

Page 11: ...t smartphones or apps When you pair with a four th client device your GlucoMen Day PENCAP will automatically drop the first client device you connected with to accommodate the new client You can clear...

Page 12: ...the GlucoMen Day PENCAP memory as well as the pairing data can be deleted by doing a Factory Reset GlucoMen Day PENCAP also reboots as a result of a factory reset 1 Unmount the Device from the pen 2 T...

Page 13: ...vice can be powered on by the following methods Connecting the Device to a charger Mounting the Device on a pen Clicking the sleeve without pen to engage the mounting mechanism 3 13 Cleaning and care...

Page 14: ...Day PENCAP radio equipment such as phones computers wireless devices antennas an tenna cables etc should not be used closer than 30 cm 12 inches from the Device This distance should also be observed...

Page 15: ...5 Unsuccessful pairing There are a number of reasons for an unsuccessful Bluetooth pairing The most common reasons are 1 It may be that the passkey was mistyped into the app or pho ne Repeat the proc...

Page 16: ...warning is displayed for a few seconds before displaying the timer When a charger is connected to the Device and the battery is not fully charged this ani mation is displayed for a few seconds before...

Page 17: ...Zed Ltd NovaUCD Belfield Innovation Park Belfied Dublin 4 Ireland By Scandinavian Healthcare Ltd Taiwan 6 3 Warranty Your GlucoMen Day PENCAP is guaranteed to be free of material and workmanship defec...

Page 18: ...function or are da maged due to obvious tampering misuse alteration neglect unau thorized maintenance or failure to operate the GlucoMen Day PENCAP in accordance with the instructions There is no othe...

Page 19: ...on IC Industry Canada Radio Module IC ID 7922A 4008 The Device meets the requirements for modular transmitter approval as detailed in RSS GEN Operation is subject to the following two condi tions 1 Th...

Page 20: ...www glucomenday com 51971 09 20 EN...

Reviews: