34
5. After drying completely, the sticky sheet can be
restored. Place it back in the carry case for use.
NOTE
The sticky sheet will become less sticky the more
you touch it, through transfer of your skin oils.
Important Control Solution Information
Use only
TaiDoc
control solution with your meter.
Do not use the control solution beyond the expiry date or 3
months after first opening. Write the opening date on your control
solution vial and discard the remaining solution after 3 months.
It is recommended that the control solution test be done at room
temperature (20°C-25°C / 68°F-77°F). Make sure your control
solution, meter and test strips are at this specified temperature
range before testing.
Shake the vial before use, discard the first drop of control solution
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Tightly close the control solution cap and store bottle between
2°C and 30°C (35.6°F and 86°F). Do NOT freeze.