36
Error Messages
MESSAGE
WHAT IT MEANS
WHAT TO DO
Appears when the
batteries cannot provide
enough power for a test.
Replace the batteries
immediately.
Appears when a used
test strip is inserted.
Repeat with a new test strip.
Problem in operation.
Repeat the test with a new
test strip.
If the meter still does
not work, please contact
GlucoRx Customer care on
01483 755133 for assistance.
Appears when test
strip is removed while
counting down, or
insufficient blood volume.
Review the instructions and
repeat test with a new test
strip. If the problem persists,
please contact GlucoRx
Customer care on 01483
755133 for help.
Appears when ambient
temperature is below
system operation range.
System operation range is
10
o
C to 40
o
C (50
o
F to 104
o
F).
Repeat the test after the
meter and test strip are in the
above temperature range.
Appears when ambient
temperature is above
system operation range.