 Replacing the battery does not affect the test results stored in the memory.

• As with all small batteries, these batteries should be kept away from small children. If  

  swallowed, promptly seek medical assistance.

• Battery might leak chemicals if unused for a long time. Remove the battery if you are 

  not going to use the device for an extended period (i.e., 3 months or more).

• Properly dispose of the battery according to your local environmental regulations.

NOTE

 Any time you wish to exit the memory, keep pressing Main button for 5 

seconds or leave it without any action for 3 minutes. The meter will switch off 
automatically.

 Control solution results are NOT included in the day average.

 If using the meter for the first time, “---“ displays when you recall the test 

results or review the average result. It indicates that there is no test result in the 
memory.

NOTE

2. Connecting to a personal computer

     Connect the cable to a USB port on your computer.   
     With the meter switched off, connect the other end of 
     the USB cable to the meter data port. "USb" will appear 
     on the meter display, indicating that the meter is in 
     communication mode.

3. Data transmission

     To transmit data, follow the instructions provided with the software. Results will be  
     transmitted with date and time. Remove the cable and the meter will automatically 
     switch off.

Your meter comes with one 3V CR2032 lithium battery. If the low battery symbol 
“     ”appears on the screen this indicates that the battery is low and it is time to replace the 
battery.

Cleaning

1. To clean the meter exterior, wipe it with one disinfecting wipe (Example:Micro-Kill Plus™) 
to clean exposed surfaces thoroughly and remove any visible dirt, blood or any other body 
fluid with the wipe. Then dry the device with a soft, dry and clean cloth.

2. Do NOT rinse the meter with water.

1. Press the edge of the battery cover and lift it up to remove.

2. Remove the old battery and replace with one 3V CR2032 lithium battery.

3. Close the battery cover. If the battery is inserted correctly, you will hear a “beep”  afterwards.

While the meter is connecting to the PC, it will be unable to perform a blood glucose 
test.

Warning: 

BATTERY RECHARGE

You can use the meter with an USB cable and the GlucoRx Healthcare Software System to 
view your test results on your personal computer. To learn more about the GlucoRx 
Healthcare Software System or to obtain a USB cable separately, please contact GlucoRx 
customer services or the place of purchase for assistance.

1. Obtaining the required cable and installing the software

     To download the GlucoRx Healthcare Software System, please visit GlucoRx website at  
     www.glucorx.co.uk.

DOWNLOADING RESULTS ONTO A COMPUTER

Distributed by GlucoRx Ltd

4 Wintonlea, Monument Way West

Woking, Surrey, GU21 5EN, UK

Website: www.glucorx.co.uk

Customer care number: 01483 755133

For self-testing.

The following is a summary of display messages. If your meter displays an error message, 
please follow the actions for the error message as described in the table below. If the 
problem persists, please contact GlucoRx for help.

DISPLAY MESSAGES AND PROBLEM-SOLVING GUIDE

MESSAGE

WHAT IT MEANS

<1.1 mmol/L (20 mg/dL)

1.1 - 33.3 mmol/L (20 - 600 mg/dL)  

>33.3 mmol/L (600 mg/dL)

STEP 4. Get result

The blood glucose result will be stored in the memory automatically.

1. Please do not change your treatment based on the result without first consulting    
    your healthcare professional.
2. Turn the meter off by removing the test strip. Discard the used test strip and lancet    
    carefully according to your local regulations.

Warning: 

STEP 1. Enter the memory mode

With the meter turned off, press Main button twice. The 
first test result will appear, indicating that you are in the 
memory mode. If you continue to press Main button, you 
can then review the last 1000 tests in the memory.

When using the meter for the first time “---” will appear, 
indicating that there are no test results in the memory.

Viewing Results 

Your Meter stores the 1000 most recent blood glucose test results with date and time in its 
memory. You can review the test results with these easy steps.

STEP 2. Recall test results

After the last test result, the most recent test result appears with the date and time. Press Main 
button to review your last 1000 test results in order. When the memory is full, the oldest 
result will be deleted when the newest is added.

STEP 3. Exit the memory mode

Press the Main button button for three seconds to turn off the meter or after the last 
result, the display will show “End” and it will turn off automatically. 

The results of control solution tests are 

NOT

 stored in the memory. Stored results are 

blood glucose results only.

NOTE

VIEWING THE METER MEMORY

For questionable results, please see test strip package insert for important information. 

Model no.: TD-4287
Dimensions: 103.4 (L) x 28.1 (W) x 15.1 (H) mm
Weight: 31.8g
Power source: 3V CR2032 lithium battery
Display: LCD
Memory: 1000 memory sets with date and time 
Automatic detection of electrode insertion
Automatic reaction time count-down
Automatic blood detection
Temperature warning
Operating conditions: 10°C to 40°C (50°F to 104°F), below 85% R.H.
Meter storage/transport conditions: -20°C to 60°C (-4°F to 140°F), below 95% R.H.
Strip storage/transport conditions: 2°C to 32°C (35.6°F to 89.6°F), below 85% R.H.
Measurement units: Fixed mmol/L
Measurement range: 1.1–33.3 mmol/L (20–600 mg/dL)

This device has been tested to meet the electrical and safety requirements of: IEC/EN 
61010-1, IEC/EN 61010-2-101, EN 61326-1, EN 61326-2-6.

SPECIFICATIONS

CARING FOR YOUR METER

SYMBOL INFORMATION

Symbol

Referent

In vitro

 diagnostic medical device

Consult instructions for use

Temperature limitation

Use by

Batch code

Manufacturer

Serial number

Authorised representative in the 

European Community

Symbol

Referent

MESSAGE

WHAT IT MEANS

ACTION

The        symbol appears with this 
message. This means that the 
battery does not have enough 
power for a test.

Replace the battery immediately.

A used test strip is inserted, or 
there is an electrical problem.

Repeat test with a new test strip. If 
the error message re-appears, 
please contact GlucoRx for help.

The temperature of the 
environment, meter, or test strip 
is outside the system’s operating 
range. You cannot perform a test 
until the meter and test strip are 
within the operating range of 
10°C to 40°C (50°F to 104°F).

Repeat the test after the meter and 
test strip are within the operating 
temperatures.

Problem with the meter.

Review the instructions and try 
again with a new test strip. If the 
problem persists, please contact 
GlucoRx for help.

You may have removed the strip 
after applying blood to the 
absorbent hole.

Keep the test strip in the meter after 
it has absorbed the blood until the 
meter displays the test result.

6

2.Press Main button to review

 14-, 21-, 28-, 60- and 90- day average results stored in each 

measuring mode in the order of Gen, AC, then PC. 

3. Exit the meter memory.

Keep pressing the Main button and the meter will turn off after displaying the last test 
result.

When         appears on the display, keep pressing Main button for 3 seconds until the 
flashing “               ” appears. Release Main button and then your 7-day average result 
measured in general mode will appear on the display. 

1. Press and release Main button.

Reviewing Blood Glucose Day Average Results