15
WARNING:
When doing the control solution test, you have to mark it so that the test
result will
NOT
mix with the blood glucose
TEST RESULTS
stored in the
memory. Failure to do so will mix up the blood glucose test results with the
control solution test results in memory.
3. Apply control solution (i)
Shake the control solution vial thoroughly before use. Squeeze out
first drop and wipe it off, then squeeze out another drop and place it
on the tip of the vial cap. Hold the meter to move the absorbent hole
of the test strip to touch the drop. Once the confirmation window fills
completely, the meter will begin counting down.
NOTICE:
To avoid contaminating the control solution, do not directly apply
control solution onto a strip.
4. Read and compare the result
After counting down to 0, the control solution test result will appear
on the display. Compare this result with the appropriate range printed
on the test strip vial, and it should fall within this range. If not, please
read the instructions again and repeat the control solution test.
NOTE:
• The control solution range printed on your test strip vial is for control
solution use only. It is not a recommended range for your pet's blood
glucose level.
• See the
MAINTENANCE
section for important information about your
control solution.