21
REPAIRS
Replacing / repairing the valve(s) solenoid – overload
protector - capacitor (s)
- Remove the cover of the connection
box (06);
- Disconnect the wires (07);
- Unscrew the cables pass (08) and
withdraw finishing spirals (09);
- Remove the wires completely.
08
09
- After loosening the motor, shift it to enable the access to the box of electric command
interior (10);
- Loosen the solenoid valve through the four screws and washers (11);
- Locate the connectors bar, remove the protection acrylic sleeve and disconnect the cables
of the motor connection (12).
06
07
10
11
12
13
14
- After loosening the cables
of the connector, remove it
pulling by the command box
orifice (13);
- Remove the motor (14).