BlueSense
GO Systemelektronik GmbH Faluner Weg 1 24109 Kiel Germany Tel.: +49 431 58080-0
Fax: -58080-11
Page 18 / 55
www.go-sys.de
info@go-sys.de
5.1.1.1 Calibration
Sensor setup
Depending on the connected sensor there is a 1-point calibration or a 2-point calibration offered.
∗
5.1.1.1.1 Example 1-point-calibration oxygen
1-Point-Calibration
Example Oxygen in mg/l:
Hold the sensor in a calibration fluid. Measure the oxygen content of the calibration fluid with a reference me-
ter. This value from the reference meter has to be entered.
Aborts the calibration and switches to Sensor control-parameter setup.
Deletes the last entered character.
Saves the input and switches to the List menu.
Without a setting the value is set to 0. Not recommended!
With the List menu you can check the calibration values.
Aborts the calibration and switches to Sensor setup.
Saves the calibration and switches to Sensor setup.
The calibration is saved and completed.
∗
Exceptions possible, see
Appendix H - Multipoint calibration
Calibration
1/1 6.2 mg/l
+/-
.
0
7
8
9
4
5
6
1
2
3
1/1
actual number of the measurement point / number of measurement points
6.2
actual measured value
mg/l
unit of the measurement
List menu
Display of the measured value
Display of the entered value (input)