GKP-2302 User Manual
86
正常時
100 V
ディップ時
70 V
周波数
: 50 Hz
5
秒
0.5
秒
5
秒
3
回繰り返し
At dipping
At normal
Frequency: 50Hz
Repeat 3 times
5 second
5 second
0.5
second
Figure 4-12 Voltage Dip Simulation Example
Table 4-6 Voltage Dip Simulation Edit Example
Step
Initial
Normal 1
Trans 1
Abnormal
Trans 2
Normal 2
Step Time
—
5 s
0 s
0.5 s
0 s
5 s
AC voltage
100 V
100 V
—
70 V
—
—
Frequency
50 Hz
50 Hz
—
50 Hz
—
—
Start phase
OFF
OFF
—
OFF
—
OFF
Stop Phase
OFF
OFF
—
OFF
—
OFF
Repeat count
2 times
Voltage change simulation
In this simulation, the 50 Hz 100 V power source repeats the voltage change three times at
10-second intervals, while the voltage drops to 70% and then linearly recovers in 0.5 seconds
during one cycle as shown in Figure 4-13. For this case, set so that the total of Normal 1 and
Normal 2 Step Times becomes 10 seconds and set the repeat count to 3. Set the Trans 1 Step
Time to zero (Table 4-7).
正常時
100 V
低下時
70 V
周波数
: 50 Hz
電圧
上昇時
5
秒
0.02
秒
0.5
秒
5
秒
3
回繰り返し
At drop
Frequency: 50Hz
Repeat 3 times
5 sec
5 sec
0.5 sec
At normal
At voltage
rise
0.02 sec
Figure 4-13 Voltage Change Simulation Example
Summary of Contents for GKP-2302
Page 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Page 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Page 60: ......
Page 186: ......
Page 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Page 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Page 216: ......
Page 222: ......