GKP-2302 User Manual
226
DC offset setting
0 mV
Measurement display mode
Normal View
Measurement unit selection
rms
Power unit energization setting
All enabled (energization)
Remote sensing
Off
AGC
Off
General function
(excluding the time unit setting of the Step
Time for the Sequence and Simulation)
Monitor output target
Current (L1 phase)
External interface
USB
External control input
Disabled
9.25
Self-diagnosis/Protection Function
At power-on
Description
ROM check
Checks the internal ROM.
RAM check
Checks the internal RAM.
Basic setting
Memory check
Checks the Basic Setting Memory.
System Setting Memory
check
Checks the System Setting Memory.
Waveform Memory
check
Checks the Waveform Memory.
Sequence Memory
check
Checks the Sequence Memory.
Simulation Memory
check
Checks the Simulation Memory.
Adjustment value
memory check
Checks the adjustment data memory.
Version check
Checks the version of the internal software.
While energizing
Description
Abnormal output
Turns off the panel display and output when an output overvoltage
or overcurrent is detected.
Power unit
error
Turns off the panel display and output when a power unit error is
detected.
Internal control error
Turns off the panel display and output when an internal
communication or other error is detected. Also stops all the
operations excluding power-off.
Summary of Contents for GKP-2302
Page 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Page 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Page 60: ......
Page 186: ......
Page 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Page 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Page 216: ......
Page 222: ......