4.1 Using the Limiter, Setting the Limiter
51
2.
Set the value of RMS current limiter in [Io] item.
3.
Set the [Output OFF] item as [Enable] or [Disable]. When setting [Output OFF] to [Enable],
also set the duration time from when the limiter is trigger to when the output is turned off.
4.
Close the window.
4.1.3
Using the Voltage Setting Range Limit
For the continuous output function, when the signal source is INT, SYNC, or ADD, the range of
the output voltage setting value to the internal signal source can be limited. Even if trying to set
a voltage that exceeds the limit, the setting value is restricted to the limiting value. This can
prevent mis-setting when you do not want to set values outside the range.
As for the voltage setting range limiting values, each value is kept independently for the
following settings for 100 V and 200 V ranges respectively: SIN/CLP waveform in AC mode,
ARB waveform in AC mode, ACDC mode, and DC mode.
The limiting value should be set using the effective value or the positive and negative peak
value. The effective value can be used only when in the AC mode and the waveform is a sine
wave or a clipped sine wave.
For the default setting range, see