GKP-2302 User Manual
56
Figure 4-1 Step-control Parameters and intra-Step Parameters
Step No. 0
Step No. 0 is assigned to the stand-by state before the sequence starts. If the terminal
specification of the step other than No. 0 is End, the execution transitions to step No. 0 and
enters into the stand-by state. Step Time and the Step Behavior setting of No. 0 specifies the
action to transit to No. 0 at the end of a sequence.
Step Behavior (Const, Keep, Sweep)
For the intra-Step parameters other than waveforms and phases, the following three patterns
can be set as the method to change values in the step.
Constant (Const)
Fixed to the value set in the step.
Keep (Keep)
Keeps the value of the current step before transitioning to the next step. That is, the value
of the step depends on the status of the previous step. Although a parameter value can be
set when the Step Behavior is set to Keep, the parameter setting value will be ignored
when the sequence is executed.
Sweep
The sweep setting changes the output value linearly from the value of the step prior to
transition to the value set in the step, taking the Step Time into consideration. In other
words, output value reaches the setting value after the Step Time has elapsed. The starting
value is depends on the status of the previous step.
ステップ動作種別
Step Behavior
Step number
Step
-
control parameters
Intra
-
Step parameters
Icon
Summary of Contents for GKP-2302
Page 15: ...1 1 OUTLINE 1 1 Overview 2 1 2 Features 2...
Page 28: ...GKP 2302 User Manual 14 Nothing is connected to the output terminal...
Page 60: ......
Page 186: ......
Page 187: ...173 5 DESCRIPTION OF SCREEN AND MENU 5 1 Screen Configuration 174 5 2 Menu Composition 177...
Page 195: ...181 6 REMOTE CONTROL 6 1 Communication Interface 182 6 2 Remote Local State Switching 188...
Page 216: ......
Page 222: ......