bAud – Setting Transmission Speed (RS 232C)
•
Specifies transmission speed in bits per second.
•
The baud rate is an operating parameter of the serial interface.
•
A low baud rate should be used in the case of large distances
between communicating devices and in EMC-critical
environments.
•
Settings at the SSP KONSTANTER and the controller must be
identical.
•
The baud rate can be set to one of the following values:
50, 75, 150, 200, 300, 600, 1200, 1800, 2400, 3600, 4800,
7200, 9600 or 19,200 bits per second.
Settings
☞
Procedure
for selecting
bAUd
, and setting the desired
transmission speed (see chapter 4.7).
!
bAUd
appears in the
bUS
display along with the related
parameter (transmission speed).
Figure 4.7.4 b
Path to Selection of the bAUd Text Parameter
dbit – Selecting the Number of Data Bits (RS 232C)
•
Either 7 or 8 data bits can be used.
•
Settings at the SSP KONSTANTER and the controller must be
identical.
Settings
☞
Procedure
for selecting
dbit
and the desired number of data bits
!
dbit
appears in the
bUS
display along with the related
parameter (number of data bits).
Figure 4.7.4 c
Path to Selection of the dbit Text Parameter
Pbit – Setting the Parity Bit (RS 232C)
•
Parity or check bits are used to detect transmission errors.
•
They are added to the character string so that the sum of H or
L is even or odd as defined.
•
The parity bit can be set to one of the following values:
none
no parity bit
ZEro
always 0
EVEn
even numbered
odd
odd numbered
nonE
always 1
!
A parity bit should always be used in EMC-critical
environments.
!
Settings at the SSP KONSTANTER and the controller must be
identical.
Settings
☞
Procedure
for selecting
Pbit
and setting the parameter (see also
!
Pbit
appears in the
bUS
display along with the related
parameter (parity bit).
Figure 4.7.4 d
Path to Selection of the Pbit Text Parameter
Select the text parameter
from an options loop
From the bUS function group
*
ENTER
CE/LOCAL
FUNCTION
SAVE
RCL
FUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change
From the bUS function group
*
ENTER
CE/LOCAL
FUNCTION
SAVE
RCL
FUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change
Select the text parameter
from an options loop
From the bUS function group
*
ENTER
CE/LOCAL
FUNCTION
SAVE
FUNCTION
MENU
SAVE
MENU
RCL
MENU
FUNCTION
MENU
Activate
the setting
Abort
without change