5
Measurement Value Storage
“DATA” (Hold & Compare)
Measurement values can be automatically “frozen” with the
DATA (Hold) function. This can be especially useful when
your full attention is required for testing the measuring
point with the test probes.
After the measurement value has been acquired, and the
appropriate “condition” has been fulfilled according to the
following table, the measurement value is displayed in the
left hand sub-display and two acoustic signal sounds. At
the same time “MAN” appears and indicates that the mea-
suring range is now set. The test probes can now be
removed from the measuring point and the measurement
value can be read from the sub-display. If the measure-
ment value lies below the limit value shown in the table, the
instrument is reactivated for the storage of a new value; the
“DATA” display blinks.
Comparison of measurement values (DATA Compare)
If the newly stored measurement value deviates less than
0.33% of the measuring range from the first measurement
value, the acoustic signal (DATA Compare) sounds twice. If
it deviates more than 0.33 % from the measuring range,
only a brief signal sounds.
DATA
Acoustic Signal
015
.
000
015
.
010
015,000