GMC-I Gossen-Metrawatt GmbH
17
GB
8
Voltage Measurement
➭
Depending upon the voltage to be measured, set the
rotary switch to V
, V
or V
.
➭
Connect the measurement cables as shown. The “
⊥
”
jack should be grounded.
Note!
☞
In the 600 V range, an intermittent acoustic signal
sounds alarm if the measurement value exceeds
1000 V.
Attention!
!
Make absolutely certain that neither of the current
ranges
(“mA” or “A”)
is active
when the multimeter is
connected for voltage measurements! If the fuse trip
limits are exceeded due to operator error, both the
operator and the instrument are in danger!
Zero Balancing in the 300 mV
Measuring Range
➭
Select the 300 mV
measuring range.
➭
Connect the measurement cables to the instrument, and
connect the free cable ends to one another.
➭
Briefly press the multifunction key
ESC|FUNC
.
015
.
000
600V
V
F
°
C
– (+)
~
+ (–)
~