23.1
Description of General Parameters in the
SEt
Menu
23.1.1 Sampling Rate:
rAtE
.
The sampling rate determines the interval, after which the
respective measurement values are transmitted to the data
interface or the measurement value memory.
The following sampling rates are possible:
0.05, 0.1, 0.2, 0.5, 00:01, 00:02, 00:05, 00:10, 00:20,
00:30, 01:00, 02:00, 05:00, 10:00, SAMPLE, dAtA.
Only METRA HIT 29S Memory mode in switch position
V
: 0.0005, 0.001, 0.002, 0.005, 0.01, 0.02.
Certain limit values apply to the various measured quanti-
ties for the given sampling rates. Actual values may not fall
below these limits (see table below).
SAMPLE (METRA HIT 29S only– Storage of Individual Values)
If the sampling rate (“rate” menu) is set to “SAMPLE”
(event), the momentary measurement value is stored to
memory by simultaneously pressing the
FUNC
and
ON
/
↵
keys. This is briefly acknowledged with
MEM
and
TRIG
in the
display and by an acoustic signal. Storage is done in the
same data block as long as the measuring function is not
changed.
dAtA
This setting provides for transmission of the measurement
values from the multimeter to the interface, or storage of
these values, which have been generated in the measure-
ment value storage function, “DATA”.
Setting the Sampling Rate
SEt
↵
V
rAtE
↵
s.zht / mm:ss
VU
↵
t: thousandths of a second, h: hundredths of a second,
z: tenths of a second, s: seconds, mm: minutes
23.1.2 Rapid Query:
Menu
CYCLE
.
After this function has been activated (MEnu
CYCLE
param-
eter set to ON) the last sub-menu to which an entry was
made is displayed, when the SET main menu is opened.
SEt
↵
V
MEnu
CYCLE
↵
OFF
VU
on
↵
23.1.3
tiME
und Datum
dAtE
Current time and date allow for real-time recording of mea-
surement values (TT: day, MM: month, JJ: year).
SEt
↵
V
tiME
↵
hh:mm:ss
hh
VU
↵
mm
VU
↵
ss
VU
↵
(hh: hours, mm: minutes, ss: seconds)
dAtE
↵
TT.MM.JJ
TT
VU
↵
MM
VU
↵
JJ
VU
↵
Measured Quantity
Sampling Rate
V
0.0005 s for memory mode
(METRA HIT 29S only)
V
, A
,
0.05 s
V
, A
,
EVENTS V
, 0.5 s
Ω, Ω
,
Count,
°
C (Pt100, Pt1000), MAinS
0.5 s
V
, Hz, dB, EVENTS V
, W, VA, VAr, Wh
1 s
°
C (J, K)
2 s
F
0.5 ... 10 s