42
GMC-I Messtechnik GmbH
V/Hz,
, Temperature,
and A/Hz Measurements
5.7.2 Alternating Current and Frequency Measurement,
Direct Connection, A AC and Hz
(METRA
HIT
X-
TRA
O
UTDOOR
T
ECH
P
RO
only)
Ð
First disconnect supply power from the measuring circuit or
the power consumer (1), and discharge any capacitors.
Ð
In accordance with the current or frequency to be measured,
turn the rotary switch to A~ or Hz.
Ð
Select the desired measured quantity by briefly pressing the
FUNC | ENTER
multifunction key. Each time the key is pressed,
AC
TRMS
and Hz are alternately selected, and
switching is acknowledged with an acoustic signal.
Ð
Safely connect the measuring instrument (without contact
resistance) in series to the power consumer as shown.
Ð
Switch supply power to the measuring circuit back on (3).
Ð
Read the display. Make a note of the measured value if the
instrument is not being operated in the memory mode or the
transmission mode.
Ð
Disconnect supply power from the measuring circuit or the
power consumer (1) once again, and discharge any
capacitors.
Ð
Remove the test probes from the measuring point and return
the measuring circuit to its normal condition.
COM
mA A
V
Temp
~
R
x
~
R
x
1
3
2
A~
Hz
A~
Hz
003
.
50
A
AC
TRMS
050
.
10
Hz
AC
8
.
8
.
8
.
8
.
8
!
I > 10 A
!
I > 16 A
Current measurement
may only be performed
with installed batteries!
Measuring Ranges:
Current
FUNC
ENTER
METRA
HIT
X-
TRA
/O
UTDOOR
METRA
HIT
P
RO
1 A / 10 A (16 A max. 30 s)
100
A / 1 mA
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
METRA
HIT
T
ECH
10 mA / 100 mA
1 A / 10 A (16 A max. 30 s)
Summary of Contents for METRAHIT Base
Page 5: ...GMC I Messtechnik GmbH 5 Contact Persons ...
Page 11: ...GMC I Messtechnik GmbH 11 Safety Precautions ...
Page 25: ...GMC I Messtechnik GmbH 25 Control Functions ...
Page 51: ...GMC I Messtechnik GmbH 51 Device and Measuring Parameters ...
Page 61: ...GMC I Messtechnik GmbH 61 Technical Data ...
Page 71: ...GMC I Messtechnik GmbH 71 ...