Safety Precautions
• Be prepared for the occurrence of unexpected voltages at
devices under test (e.g. defective devices). For example,
capacitors may be dangerously charged.
• Make certain that the measurement cables are in flawless
condition, e.g. no damage to insulation, no interruptions in
cables or plugs etc.
• No measurements may be made with this instrument in
electrical circuits with corona discharge (high-voltage).
• Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be
present.
• Measurements under moist ambient conditions are not
permitted.
• Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8, “Technical Data”, in the table entitled
“Measuring Functions and Measuring Ranges” in the “Over-
load Capacity” column.
• The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents and voltages are
otherwise not indicated, and the instrument may be damaged.
• The instrument may not be operated if the fuse cover or the
battery compartment lid has been removed, or if its housing is
open.
• The input for the current measuring range is equipped with a
fuse link (exception:
METRA
HIT
B
ASE
).
Maximum permissible voltage for the measuring circuit (= rated
voltage of the fuse) is 1000 V AC/DC.
Use specified fuses only (see page 59)! The fuse must have a
breaking capacity
of at least 30 kA.
Repair and Parts Replacement
When the instrument is opened, voltage conducting parts may be
exposed. The instrument must be disconnected from the measur-
ing circuit before the performance of repairs or the replacement of
parts. If repair of a live open instrument is required, it may only be
carried out by trained personnel who are familiar with the dangers
involved.
Defects and Extraordinary Strains
If it may be assumed that the instrument can no longer be
operated safely, it must be removed from service and secured
against unintentional use.
Safe operation can no longer be relied upon:
• If the device demonstrates visible damage
• If the instrument no longer functions, or if malfunctioning
occurs
• After long periods of storage under unfavorable conditions,
e.g. humidity, dust or extreme temperature (see “Ambient Con-
ditions” on page 60)
Summary of Contents for METRAHIT Base
Page 5: ...GMC I Messtechnik GmbH 5 Contact Persons ...
Page 11: ...GMC I Messtechnik GmbH 11 Safety Precautions ...
Page 25: ...GMC I Messtechnik GmbH 25 Control Functions ...
Page 51: ...GMC I Messtechnik GmbH 51 Device and Measuring Parameters ...
Page 61: ...GMC I Messtechnik GmbH 61 Technical Data ...
Page 71: ...GMC I Messtechnik GmbH 71 ...