46
GMC-I Messtechnik GmbH
V/Hz,
, Temperature,
and A/Hz Measurements
5.7.5 Direct, Pulsating and Current Measurement with Current Clamps
Transformer A DC, A (DC+AC), A AC and Hz (METRAHIT E
XTRA
,
METRAHIT E
TECH
and METRAHIT E
SPECIAL
only)
Transformer Output, Current/Current
When a current clamps transformer is connected to the multime-
ter ( mA/A input), all current displays appear with the correct
value in accordance with the selected transformation factor. The
only prerequisite is that the current transformer is equipped with
at least one of the below listed transformation factors, and that
the factor has been previously selected in the following menu (
Cl
ip
0FF
), see also chapter 6.4.
Current Clamp Setup Menu
1nFo
...
SET
rAtE
...
CLIP
1
/
10
/
100
/
1000
/
0ff
(default setting:
Cl
ip
0FF
)
Trans. Factors
CL
IP
Measuring Ranges DMM
Clamp Types
60 mA AC
600 mA AC
6 A AC
1:1
1mA/1mA
60.000 mA
600.00 mA
6.0000 A
WZ12A, WZ12D,
WZ11A, Z3511,
Z3512, Z3514
1:10
1mA/10mA
600.00 mA
6.0000 A
60.000 A
1:100
1mA/100mA
6.0000 A
60.000 A
600.00 A
1:1000
1 mA/1 A
60.000 A
600.00 A
6000.0 A
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM
V
Temp
A
~
R
x
Current
1:1000
1:1000
R
i
R
i
1
mA
A~
Hz
A~
Hz
003
.
50
A
AC
TRMS
050
.
10
Hz
AC
FUNC
ENTER
1:1000
1:1000
A~
Hz
A~
Hz
003
.
50
A
AC
TRMS
050
.
10
Hz
AC
FUNC
ENTER