45
Gossen Metrawatt GmbH
Device and Measuring Parameters
8.3 Querying Parameters – InFo Menu
(as moving letters)
bAtt
–
query battery voltage
1nFo
bAtt
:
2
.
75
V.
vErSion – query firmware version
1nFo
bAtt
:
vErSion
:
2
.
09
ItEMP – query reference temperature
The temperature of the internal reference junction is measured with a
temperature sensor in close proximity to the input sockets.
1nFo
bAtt
:
...
1tEMP
:
24
°C
dAtE – query date
1nFo
bAtt
:
...
dAtE
:
31
.
12
.
05
(DD.MM.YY)
D = day, M = month, Y = year
Date and time must be reentered after replacing the batteries.
tiME – query time
1nFo
bAtt
:
...
t
iME
:
13
:
46
:
56
(hh:mm:ss)
h = hours, m = minutes, s = seconds
Date and time must be reentered after replacing the batteries.
OCCUP
–
query memory occupancy (METRAHIT PM XTRA / PM TUNE /
E
XTRA
only)
1nFo
bAtt
:
...
0CCUP
:
000
.
0
%
8.4 Entering Parameters – SETUP Menu
rAtE
–
set the sampling rate (METRAHIT PM XTRA / PM TUNE / E
XTRA
only)
The sampling rate specifies the time interval after which the
respective measured value is transmitted to the interface, or to
measured value memory.
Any one of the following rates can be selected:
00:00.1, 00:00.2,
00:00.5
, 00:01.0, 00:02.0, 00:05.0
[h:mm:ss.t] (h = hours, m = minutes, s = sec., t = tenths of a sec.)
0:00:10, 0:00:20, 0:00:30, 0:00:40, 0:00:50, 0:01:00, 0:02:00, 0:05:00,
0:10:00, 0:20:00, 0:30:00, 0:40:00, 0:50:00, 1:00:00, 2:00:00, 3:00:00,
4:00:00, 5:00:00, 6:00:00, 7:00:00, 8:00:00, 9:00:00
Setting the Sampling Rate
1nFo
...
SET
rAtE
00
:
00
.
1
...
00
:
00
.
5
...
9
:
00
:
00
(00:00.5 = 0.5 seconds = default value)
The last value is retained.
noFiL – quick display of measured values
For the following functions, a quicker display update rate of 5 dis
-
play values per second (
noFiL = ON
) instead of 2 display values per
second (
noFiL = OFF
) can be set: V DC, A DC with clamp, A DC direct,
, diode measurement, continuity test.
1nFo
...
SET
rAtE
...
nofil
on
/
0ff
(standard value/default setting for
noFil
: PM TUNE
ON
, all others
OFF.
)
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER