63
Gossen Metrawatt GmbH
Accessories
12 Accessories
12.1 General
The extensive accessories available for our measuring
instruments are checked for compliance with currently valid safety
regulations at regular intervals, and are expanded as required for
new applications. Currently up-to-date accessories which are
suitable for your measuring instrument are listed at the following
web address along with photo, order number, description and,
depending upon the scope of the respective accessory, data
sheet and operating instructions: www.gossenmetrawatt.com.
12.2 Technical Data for Measurement Cables
(included: KS17-2 safety cable set)
Electrical Safety
* Exception
METRAHIT EU PRO /
E
SPECIAL
: 600 V
Ambient Conditions (EN 61010-031)
Temperature
–20
C ... + 50
C
Relative humidity
max. 80%
Pollution degree
2
Application KS17-2
Attention!
!
In conformity with standard DIN EN 61010-031, mea
-
surements in an environment according to measuring
category III and IV may only be performed with the safety
cap applied to the test probe of the measurement cable
(does not apply to
METRAHIT EU PRO /
E
SPECIAL
).
For establishing contact in 4 mm jacks you have to remove the
safety cap by levering out the snap lock of the safety cap with an
-
other sharp object (e.g. the second test probe).
12.3 NA X-TRA Power Pack (not included)
Use only the power pack from Gossen Metrawatt GmbH in com
-
bination with your instrument. This assures operator safety by
means of an extremely well insulated cable, and safe electrical
isolation (nominal secondary ratings: 5 V / 600 mA). Installed
batteries are disconnected electronically if the power pack is
used, and need not be removed from the instrument.
Note
Use of the power pack may result in an additional mea
-
suring error on account of capacitive
coupling. We there
-
fore recommend battery operation for measuring capaci
-
tance and AC quantities.
The specified technical data only apply for battery opera
-
tion.
Maximum Rated Voltage
600 V
1000 V 1000 V*
Measuring Category
CAT IV
CAT III CAT II
Maximum Rated Current
1 A
1 A
16 A
with safety cap applied
—
without safety cap applied
—
—