34
GMC-I Messtechnik GmbH
6.10
Temperature Measurement with Thermocouple – Temp
TC
Temperature measurement is performed with a type K thermo-
couple (accessory, not included), which is connected to the volt-
age input.
➭
Set the rotary switch to “
” or “Temp”.
➭
Repeatedly press the
Func
softkey until the
Temp TC
measuring
function is displayed.
The reference temperature is measured via an internal reference
junction. It’s displayed as T
INT
, or it can be queried via “General
Setup” (see description below).
The “Setup for currently selected measurement” menu can be
used to specify whether or not the internal reference junction tem-
perature or a manually entered temperature will be used (see
below). If “Manual” temperature is selected, T
MAN
appears that
the display.
➭
Connect the sensor to the sockets with the measurement ca-
bles as shown.
➭
Measurement is started immediately. The instrument displays
the measured temperature using the selected unit of measure.
Querying the Measured Reference Temperature
> General Setup > Info >
Temperature xx.x °C
Note
The internal reference temperature (temperature of the
internal reference junction) is measured by a temperature
sensor inside the instrument. This may be somewhat
higher or lower than room temperature as a result of inter-
nal heat-up, or moving from warmer to colder surround-
ings or vice versa.
Selecting the Unit of Measure for Temperature
> Setup for currently selected measurement
(
Temp
TC
Setup
)
Temperature U/M
°C / °F
(
°C
= default setting)
Entering “Manual” Reference Temperature Tman
> Setup for currently selected measurement
(
Temp
TC
Setup
)
Tman
+xx.x °C
Choice Between Measured and “Manual” Reference Temperature
> Setup for currently selected measurement
(
Temp
TC
Setup
)
Compensation Type
Man / Int
Man
Manually specified reference temperature
Int
Internally measured reference temperature
MENU
MENU
OK
OK
MENU
OK
OK
MENU
OK
OK
S–
A
V
Temp
m
/4
COIL
S+
Temp
TC
Measuring Range
TC
K (NiCr-Ni) –250.0
+ 1372.0
C
Func
Func
Func
Func
F
Temp. RTD
TEMP TC