Measurements: V/Hz,
,
, Temp (°C/°F),
, M
@UISO
, A/Hz
5.8.4 Alternating Current Measurement with Clip-On Current Sensor,
A AC and Hz
Transformer Output, Voltage/Current
When a clip-on current sensor is connected to the multimeter
(V input), all current displays appear with the correct value in
accordance with the selected transformation ratio. The only
prerequisite is that the current sensor is equipped with at least
one of the below listed transformation ratios, and that the ratio
has been previously selected in the following menu (
Cl
ip
0FF
).
Current Clip Setup Menu
1nFo
...
SET
rAtE
...
CL
IP
0ff
/
1
/
10
/
100
/
1000
Maximum allowable operating voltage is equal to the current
transformer’s nominal voltage. When reading the measured value,
additional error resulting from the clip-on current sensor must also be
taken into consideration.
(default value:
Cl
ip
=
0ff
= voltage display)
Transf. Ratio
CL
IP
Measuring Ranges
Clip Type
300 mV
3 V
30 V
1:1
1 mV / 1 mA
300.0 mA
3.000 A
30.00 A
WZ12C
1:10
1m V / 10 mA
3.000 A
30.00 A
300.0 A
WZ12B,
Z201A/B
1:100
1m V / 100 mA
30.00 A
300.0 A
3,000 kA
Z202A/B
METRAFLEX
1:1000
1 mV/1 A
300.0 A
3,000 kA
30.00 kA
Z202A/B, Z203A/B,
WZ12C
METRAFLEX
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Temp
M
ISO
0
035
.
0
A
AC
TRMS
A~
Hz
1:1000
0
050
.
1
Hz
AC
1:1000
FUNC
ENTER
V~
Hz
~
R
x
Current
R
i
~ 9 M
R
i
Measuring Ranges:
Hz:
1 Hz . . . 31 kHz
MEASURE
SETUP
...
SET
...
CL
IP
FUNC
ENTER
FUNC
ENTER
0ff
mA / A:
see table
3 ranges:
300 Hz, 3 kHz and 30 kHz