Troubleshooting
28
3A7649A
Disinfectant leaks out of
the cup threads.
Cup not properly seated.
Make certain that there is only
one FlexLiner in Cup Support.
Make certain the Cup Lid is
properly threaded to the Cup
Support. If threads are visible
below the Cup Support when
tight, fully remove the Cup Lid
and reinstall to the Cup Support
so no threads are visible when
tight.
Make certain the Cup Lid is
tightened to the Cup Support so
the arrow on the Cup Lid is within
range of the indicator on the Cup
Support.
Avoid flexing or pushing on the
Cup Support when you evacuate
the air out of the FlexLiner.
Avoid pulling down on the
FlexLiner when you evacuate the
air out of the FlexLiner.
Make certain there is no damage
to the FlexLiner lip or the Cup Lid
gasket.
Make certain the FlexLiner lip
and Cup Lid gasket are free of
debris.
Replace FlexLiner.
Problem
Cause
Solution